Cardiac resynchronization therapy pacemaker

Advertisement

Patient follow‑up visits should be scheduled at an appropriate frequency so that ERI can be detected well
before end‑of‑life (EOL).
Clearing ERI
When the programmer displays a message that the device has reached ERI, you are able to clear ERI. For
additional information on clearing ERI, refer to the programmer's on‑screen help.
CAUTION: Programming to high output settings or a high Base Rate may shorten the time to ERI.
Programming to lower rates and outputs may restore normal battery status.
If the programmer displays an ERI warning message, the clinician should fully evaluate the device.
WARNING: At ERI, the nominal life of the device is six months. When the device exhibits signs of ERI
(described on the programmer's on‑screen help), it should be replaced expeditiously.
End-of-Life
End‑of‑life (EOL) occurs when the battery voltage has fallen to a level indicated below.
The approximate EOL battery voltage for all devices in 2.47V.
For additional information, refer to the programmer's on‑screen help.
Explanting the Pulse Generator
WARNING: Before explanting the system or disconnecting the leads from a pulse generator, program the
pulse generator to disable all pacing. In the event of the patient's death, deactivate the pulse generator
before post‑mortem examination.
Explant the device with standard surgical tools.
If a lead or adapter is explanted, be careful not to damage it during removal.
37

Advertisement

loading
Need help?

Need help?

Do you have a question about the Endurity Core and is the answer not in the manual?

Questions and answers

Tori Garvin
January 25, 2025

My mother has model number PM2272 cardiac pacemaker. Can she have MRI Done. She has been in hospital for 3 days. Doctors want do a MRI until they hear from manufactory. Makes no since. They have her card info.

User image 67951fc7942f7
3 comments:
HK
February 8, 2025

@Tori Garvin, of couse 1.5T and 3T

Mr. Anderson
February 10, 2025

The Abbott Endurity Core PM2272 is not listed among the MR Conditional devices in the provided context. Before performing an MRI scan on a patient with this pacemaker, it is necessary to contact Abbott Medical or consult regulatory authorities to determine if it is approved for MRI use.

This answer is automatically generated

HK
February 11, 2025

@Mr. Anderson PM2272 isn't Endurity Core, It's Assurity MRI; of couse 1.5T and 3T MRI Scan

Kristina Gardner
April 15, 2025

Abbott pacemaker PM2272, is it safe for Hyperbaric oxygen therapy? If so, at what ATA- 2 or 2.5 safe? Please email the information to me if possible. Thank you!

Aldo De Giuli
May 14, 2025

Is Abbott Endurity 2162 Pace Maker MRI compatible ?

James Sean
January 20, 2025

can batteries be replaced in your model no. PM2272 pacemaker

1 comments:
Mr. Anderson
May 14, 2025

No, the battery in the Abbott Core PM2272 pacemaker cannot be replaced separately. When the battery reaches the end of its lifespan, the entire pacemaker must be replaced.

This answer is automatically generated