Ultrasound Treatment - Abbott Endurity Core User Manual

Cardiac resynchronization therapy pacemaker

Advertisement

Implantation
Ensure that a separate standby external defibrillator is immediately available.
Diathermy
Avoid diathermy, even if the device is programmed off, as it may damage tissue around the implanted
electrodes or may permanently damage the pulse generator.
Electrosurgery
Do not use electrosurgical devices in the vicinity of an implanted device. If electrocautery is necessary,
use a bipolar cauterizer or place the indifferent electrode as far from the device as possible.
Lithotripsy
Avoid lithotripsy unless the therapy site is not near the pulse generator and leads as lithotripsy may
damage the pulse generator.
Therapeutic Radiation
Do not use therapeutic ionizing radiation in the vicinity of an implanted device. Radiation therapy may
damage the electronic circuitry of the device. Any damage to the device by therapeutic ionizing radiation
may not be immediately detectable.

Ultrasound Treatment

The device should not be exposed to therapeutic levels of ultrasound energy, as the device can
inadvertently concentrate the ultrasound field and cause harm that might not be immediately detectable.
Diagnostic ultrasound treatment is not known to affect the function of the device.
9

Advertisement

loading
Need help?

Need help?

Do you have a question about the Endurity Core and is the answer not in the manual?

Questions and answers

Tori Garvin
January 25, 2025

My mother has model number PM2272 cardiac pacemaker. Can she have MRI Done. She has been in hospital for 3 days. Doctors want do a MRI until they hear from manufactory. Makes no since. They have her card info.

User image 67951fc7942f7
3 comments:
HK
February 8, 2025

@Tori Garvin, of couse 1.5T and 3T

Mr. Anderson
February 10, 2025

The Abbott Endurity Core PM2272 is not listed among the MR Conditional devices in the provided context. Before performing an MRI scan on a patient with this pacemaker, it is necessary to contact Abbott Medical or consult regulatory authorities to determine if it is approved for MRI use.

This answer is automatically generated

HK
February 11, 2025

@Mr. Anderson PM2272 isn't Endurity Core, It's Assurity MRI; of couse 1.5T and 3T MRI Scan

Kristina Gardner
April 15, 2025

Abbott pacemaker PM2272, is it safe for Hyperbaric oxygen therapy? If so, at what ATA- 2 or 2.5 safe? Please email the information to me if possible. Thank you!

James Sean
January 20, 2025

can batteries be replaced in your model no. PM2272 pacemaker