Ventricular Sensing
▪
▪
In CRT‑Ps, Ventricular Sensitivity should be programmed to the highest setting (lowest sensitivity) that will
provide ventricular sensing with adequate sensing margin. Left ventricular lead dislodgement to a
position near the atria can result in atrial oversensing and ventricular inhibition.
▪
▪
Perform a thorough assessment of device function following exposure to any of the above.
Magnetic Resonance Imaging (MRI)
▪
▪
MR Conditional pulse generators and CRT‑Ps. Testing has demonstrated that the Abbott Medical
MR Conditional system is conditionally safe for use in the MRI environment when used according to the
instructions in the MRI‑Ready Systems manual. The Abbott Medical MR Conditional system includes an
Abbott Medical MR Conditional device connected to one or more Abbott Medical MR Conditional leads.
▪
▪
MRI untested pulse generators and CRT‑Ps. "Untested" indicates that the device has not been tested, and
its use in an MRI environment is not determined.
▪
▪
Scanning patients who have other MR Conditional devices that are not implanted in cardiac tissue is
acceptable provided all MR Conditional requirements for each implanted device are met.
Backup VVI
▪
▪
In rare instances, the device may revert to Backup VVI operation at the settings listed in the table below.
These values are not programmable.
▪
▪
When the device has reverted to Backup VVI operation, the programmer displays a pop‑up message
indicating that the device is operating at the Backup VVI values. Press the Continue button and follow the
on‑screen instructions.
10
Need help?
Do you have a question about the Endurity Core and is the answer not in the manual?
Questions and answers
My mother has model number PM2272 cardiac pacemaker. Can she have MRI Done. She has been in hospital for 3 days. Doctors want do a MRI until they hear from manufactory. Makes no since. They have her card info.
@Tori Garvin, of couse 1.5T and 3T
The Abbott Endurity Core PM2272 is not listed among the MR Conditional devices in the provided context. Before performing an MRI scan on a patient with this pacemaker, it is necessary to contact Abbott Medical or consult regulatory authorities to determine if it is approved for MRI use.
This answer is automatically generated
@Mr. Anderson PM2272 isn't Endurity Core, It's Assurity MRI; of couse 1.5T and 3T MRI Scan
Abbott pacemaker PM2272, is it safe for Hyperbaric oxygen therapy? If so, at what ATA- 2 or 2.5 safe? Please email the information to me if possible. Thank you!
can batteries be replaced in your model no. PM2272 pacemaker