Table 8. Expected device lifetime
Model name
Assurity MRI
Zenex MRI
Allure RF
Quadra Allure MP RF
Quadra Allure
Quadra Allure MP
NOTE: For SR models, VVI is the most comprehensive pacing mode.
NOTE: For DR models, DDD is the most comprehensive pacing mode.
NOTE: For CRT‑P models, DDD Bi‑V is the most comprehensive pacing mode.
Elective Replacement Indicator (ERI)
ERI (or Recommended Replacement Time) is the point at which battery voltage has dropped to the lowest
capacity that will maintain adequate pulse generator operation for a nominal period of before end of life
(EOL).
The time between ERI and EOL for all devices is 6 months.
When the device reaches ERI, a number of indicators alert the clinician to this condition. For information on
these conditions, refer to the programmer's on‑screen help.
Model number
PM2272
PM2282
PM3222
PM3262
PM3542
PM3562
Device type
DR
DR
CRT‑P
CRT‑P
CRT‑P
CRT‑P
36
Expected device lifetime
9.9 years
9.9 years
8.3 years
7 years
8.6 years
7.3 years