■ Characteristics
Display range
Measurement range
Input signals
Connectable sensors
Max. No. of display digits
Comparative output response
time
Linear output response time
Insulation resistance
Dielectric strength
Noise immunity
Vibration resistance
Shock resistance
Weight
Degree of
Front panel Conforms to NEMA 4X for indoor use (equivalent to IP66)
protection
Rear case
Terminals
Memory protection
Applicable standards
EMC
19,999 to 99,999
Functions F1, F2: ±2 gigacounts
Functions F3
: 0 to 4 gigacounts
• No-voltage contact (30 Hz max. with ON/OFF pulse width of 15 ms min.)
• Voltage pulse
Mode
Input frequency
F1
0 to 30 kHz
F2
0 to 25 kHz
F3
0 to 50 kHz
• Open collector
Mode
Input frequency
F1
0 to 30 kHz
F2
0 to 25 kHz
F3
0 to 50 kHz
ON residual voltage: 3 V max.
OFF leakage current: 1.5 mA max.
Load current:
Must have a switching capacity of 20 mA or higher.
Must be able to properly switch load currents of 5 mA or less.
5 ( 19999 to 99999)
1 ms max.: Transistor output; 10 ms max.: Relay contact output
(time until the comparative output is made when there is a forced sudden change in the input signal from 15%
to 95% or 95% to 15%)
10 ms max. (time until the final analog output value is reached when there is a forced sudden change in the
input signal from 15% to 95% or 95% to 15%)
20 M min. (at 500 VDC)
2,300 VAC for 1 min between external terminals and case
100 to 240 VAC models:
1,500 V at power supply terminals in normal or common mode
(waveform with 1-ns rising edge and pulse width of 1 s/100 ns)
24 VAC/VDC models:
1,500 V at power supply terminals in normal or common mode
(waveform with 1-ns rising edge and pulse width of 1 s/100 ns)
Frequency: 10 to 55 Hz; Acceleration: 50 m/s
2
2
150 m/s
(100 m/s
for relay outputs) 3 times each in 3 axes, 6 directions
Approx. 300 g (Base Unit only)
IP20
IP00 + finger protection (VDE0106/100)
EEPROM (non-volatile memory)
Number of rewrites: 100,000
UL61010C-1, CSA C22.2 No. 1010.1 (evaluated by UL)
EN61010-1 (IEC61010-1): Pollution degree 2/Overvoltage category II
EN61326: 1997, A1: 1998, A2: 2001
EMI: EN61326+A1 industrial applications
Electromagnetic radiation interference
CISPR 11 Group 1, Class A: CISPRL16-1/-2
Terminal interference voltage
CISPR 11 Group 1, Class A: CISPRL16-1/-2
EMS: EN61326+A1 industrial applications
Electrostatic Discharge Immunity
EN61000-4-2: 4 kV (contact), 8 kV (in air)
Radiated Electromagnetic Field Immunity
EN61000-4-3: 10 V/m 1 kHz sine wave amplitude modulation (80 MHz to 1 GHz, 1.4 to 2 GHz)
Electrical Fast Transient/Burst Immunity
EN61000-4-4: 2 kV (power line), 1 kV (I/O signal line)
Surge Immunity
EN61000-4-5: 1 kV with line (power line), 2 kV with ground (power line)
Conducted Disturbance Immunity
EN61000-4-6: 3 V (0.15 to 80 MHz)
Power Frequency Magnetic Immunity
EN61000-4-8: 30 A/m (50 Hz) continuous time
Voltage Dips and Interruptions Immunity
EN61000-4-11: 0.5 cycle, 0 /180 100% (rated voltage)
ON/OFF
ON voltage
range
pulse width
16 s min.
4.5 to 30 V
20 s min.
9 s min.
ON/OFF
range
pulse width
16 s min.
20 s min.
9 s min.
2
, 10 sweeps of 5 min each in X, Y, and Z directions
Up/Down Counting Pulse Indicator
OFF voltage
impedance
30 to 2 V
10 k
Note: The Up/Down Counting Pulse
Meter will malfunction if a pulse
greater than the input frequency
range is input. SYSERR may
appear on the display.
K3HB-C
Input
17

Advertisement

Table of Contents
loading

This manual is also suitable for:

K3hb-rK3hb-pK3hb-c

Table of Contents