Inficon IRwin S Original Operating Instructions

Inficon IRwin S Original Operating Instructions

Portable methane leak detector
Hide thumbs Also See for IRwin S:
Table of Contents

Advertisement

Quick Links

Original operating instructions
®
IRwin
S/SX/SXT/SXG/SXGT
Portable Methane Leak Detector
Catalog No.
580-000, 580-010, 580-015, 580-020, 580-030
From software version
4.06.01
mina66en1-21-(2403)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IRwin S and is the answer not in the manual?

Questions and answers

Summary of Contents for Inficon IRwin S

  • Page 1 Original operating instructions ® IRwin S/SX/SXT/SXG/SXGT Portable Methane Leak Detector Catalog No. 580-000, 580-010, 580-015, 580-020, 580-030 mina66en1-21-(2403) From software version 4.06.01...
  • Page 2 INFICON AB Wahlbecksgatan 25A SE-582 13 Linköping Sweden...
  • Page 3: Table Of Contents

    INFICON Table of Contents Table of Contents 1 About these instructions .......................... 6 1.1 Target groups............................ 6 1.2 Warnings .............................. 6 1.3 Patents .............................. 6 2 Safety................................ 7 2.1 Intended use ............................ 7 2.2 Special conditions for safe use ........................  8 2.2.1 Intrinsic Safety (Ex protection) ......................
  • Page 4 Table of Contents INFICON 5.2.6 Bluetooth Pairing Code required.................... 46 5.2.7 Change permission level...................... 46 5.2.8 PPM Adjustment Factor ...................... 47 5.2.9 Select Gas (Ex)...........................  47 5.2.10 100% CH4 LEL conc........................ 47 5.2.11 Set limits and alarm levels ...................... 48 5.2.12 Select Available Modes.......................
  • Page 5 INFICON Table of Contents 5.3.11.3 Performing IR Ethane Analysis after automatic reminder.......... 79 5.3.11.4 Starting IR Ethane Analysis manually................ 80 5.3.11.5 Viewing IR Ethane Analysis info .................. 80 5.3.12 Calibration (adjustment)...................... 81 5.3.12.1 Standard calibration procedures.................. 84 5.3.12.2 Setting of calibration gas concentrations ................ 85 5.3.12.3 Adjust calibration reminder ....................
  • Page 6: About These Instructions

    1 | About these instructions INFICON 1 About these instructions This document applies to the software version stated on the title page. Product names may occur in the document, which are added for identification purposes only and belong to the respective owner of the rights.
  • Page 7: Safety

    EX application it is certified for (see "EX certification (intrinsic safety) [} 23]"). These models are jointly referred to as IRwin SXnn in the following. IRwin S and accessories other than the included, original INFICON hand probe are not EX certified.
  • Page 8: Special Conditions For Safe Use

    The charger or car adapter is subsequently connected to the inlet of the Ex Certified IRwin Charging Adapter. This means that the charger for IRwin S must, under no circumstances, be connected directly to the charging socket on any of IRwin models SX, SXT, SXG, and SXGT. See also Charging the battery [} 37].
  • Page 9 INFICON Safety | 2 IRwin SXnn is certified for use in Gas groups IIA, IIB and IIC (excluding Acetylene) and in Zones 0, 1 and 2 "Declarations of conformity [} 106]". WARNING IRwin must not be used in applications where there is risk for exposure to acetylene.
  • Page 10: Certified Gas Measurement

    (called Calibration in this manual), see "Setting the Function Test Interval [} 51]" and ”Calibration (adjustment) [} 81]”. IRwin S, IRwin SX and IRwin SXG are operable within -20 to +50°C. The certification is valid in an ambient temperature of -15 to +40°C. The accuracy may not be according to specification outside this range.
  • Page 11: Owner/Supervisor Responsibilities

    (yellow box) has been performed by individuals or organizations not qualified therefore in writing by INFICON AB, Sweden. The manufacturer (INFICON) denies all responsibility for the compliance of this product with any of the type certificates for this product if the equipment is used in any way not conforming to the instructions in this User Manual.
  • Page 12: Duties Of The Operator

    2 | Safety INFICON 2.4 Duties of the operator • Read, observe, and follow the information in this manual and in the work instructions provided by the owner. This concerns in particular the safety instructions and warnings. • Always observe the complete operating instructions for all work.
  • Page 13: Description

    • Flexi Bell is used for simple survey including meters and piping. The bell adapts to pipe geometry. INFICON single-cell IRwin methane leak detector uses the unique INFICON single-cell wide range wide range technology for detecting all concentrations from 1 ppm to 100 Vol% methane. This technology (patent technology is very sensitive to the ethane content of natural gas.
  • Page 14: Design Of The Leak Detector

    3 | Description INFICON preferred you can instead set the ethane content manually. The safety certified modes Ex and Ex Tox are not affected by this phenomenon. See "IR Ethane Analysis and compensation [} 77]". IRwin SXG and SXGT models also have GC Ethane Analysis capability for distinguishing between Natural gas (NG), liquid petroleum gas (LPG) and swamp gas by determining the presence of methane, ethane and propane in the sampled gas.
  • Page 15: Probes

    WARNING Risk of explosion due to not approved probe parts. The original INFICON Hand probe (part number 580-100) is the only part of the probe system that is certified for use in potentially explosive areas. Other parts of the probe system are not allowed in classified hazardous areas (Zone 0, 1, 2, Div 1 or 2).
  • Page 16 3 | Description INFICON Hand probe You can use the INFICON Hand probe to search for gas in building installations. The hand probe also constitutes the handle for the INFICON Carpet probe and for the INFICON Bell probe. Fig. 4: Hand probe...
  • Page 17 INFICON Description | 3 Swan Neck The INFICON Swan Neck is used on uneven surfaces where the carpet probe is unpractical. Can also be practical in bushes and other “confined” areas. Fig. 7: Swan Neck Flexi Bell The INFICON Flexi Bell is practical for surveys including frequent entries into gardens and check of meters and piping.
  • Page 18: Display

    3 | Description INFICON 3.4 Display (10 PPM) Above ground Fig. 9: Display IRwin Navigation buttons Navigation and on / off button Battery status indicator Mode icon Measured value, target gas and Measurement range (in % or PPM unit depending on the operation...
  • Page 19 INFICON Description | 3 Function buttons OK/Select/Confirm Navigation buttons (in this case go right) Cancel process Start process Start data logging Stop data logging Mute the alarm Table 2: Function buttons Symbols of the status ® Bluetooth activated Symbol ON = GPS connected and...
  • Page 20 3 | Description INFICON Permission levels Symbol Permission level Default password Basic none Intermediate 1111 Advanced (Supervisor) 1422 Table 4: Permission levels Status LED Green • Leak detector is charging normal (power off) • Normal operation in the modes Ex and Ex Tox.
  • Page 21: Technical Data

    Description | 3 3.5 Technical data Mechanical data 580-000 580-010 580-015 580-020 580-030 Name IRwin S IRwin SX IRwin SXT IRwin SXG IRwin SXGT Weight 1.4 kg (3 lb.) 1.6 kg (3.5 lb.) 1.6 kg (3.5 lb.) 1.6 kg (3.5 lb.) 1.6 kg (3.5 lb.)
  • Page 22 3 | Description INFICON Physical data 580-000 580-010 580-015 580-020 580-030 Name IRwin S IRwin SX IRwin SXT IRwin SXG IRwin SXGT Noise level < 70 dB (A) < 70 dB (A) < 70 dB (A) < 70 dB (A) <...
  • Page 23: Ex Certification (Intrinsic Safety)

    • IRwin SX 580-010, • IRwin SXT 580-015, • IRwin SXG580-020, • IRwin SXGT580-030, are Ex certified with rating according to the following table. IRwin S 580-000 is not Ex certified. EX rating EX rating Ex ia IIC T3, T : -20°C - +50°C,...
  • Page 24: Sensors

    3 | Description INFICON 3.5.2 Sensors Basic sensor data Gas (measurement range) Sensor Warm up time Infrared (IR) < 30 s Infrared (IR) < 30 s , CO, H Electrochemical < 120 s sensor: 1 - 12 hours if stored with flat battery.
  • Page 25 INFICON Description | 3 Universal mode This mode is available if Universal Mode option is selected, See Operating modes Universal and Universal+ [} 56]. Measurement Infrared (IR) principle Measurement : 1 ppm - 100 Vol% range The following is also displayed when total NG is above 2.7 Vol%: content in NG: 0.5 - 8 Vol%...
  • Page 26 3 | Description INFICON Universal+ mode This mode is available if Universal+ option is selected, see "Operating modes Universal and Universal+ [} 56]". Measurement Infrared (IR) principle Measurement : 1 ppm - 100 Vol%* range % LEL displayed in three ranges**: <...
  • Page 27 INFICON Description | 3 This value is not compliant with or certified to EN50104 and must not be used for safety clearance, see "Certified Gas Measurement [} 10]": # Reduced number of gases can be used, see "Concentration of calibration and test gases [} 35]"...
  • Page 28 3 | Description INFICON GC Ethane Analysis mode: Measurement Gas Chromatography separation with semiconductor detector principle Measurement 1000 ppm gas needed in sample. This refers to total range hydrocarbon content (i.e., natural gas, swamp gas or LPG etc.). Capability Can identify Natural Gas at 0.5% C level.
  • Page 29 INFICON Description | 3 House mode: This mode is not available if Universal mode option is activated. Measurement Infrared (IR) principle Measurement : 1 ppm - 100 Vol% range Resolution 1 ppm (0 - 100 ppm), 10 ppm (110 - 990 ppm), 0.1 Vol% (0.1 - 1 Vol%), 1 Vol% (>5 Vol%)
  • Page 30 3 | Description INFICON Plants mode: This mode is not available if Universal mode option is activated. Measurement Infrared (IR) principle Measurement : 1 ppm - 5 Vol% range Resolution 1 ppm (0 - 100 ppm), 10 ppm (110 - 990 ppm), 0,1 Vol% (0,1 - 5...
  • Page 31 INFICON Description | 3 Ex mode: This mode is not available in IRwin S. Measurement Infrared (IR) principle Measurement : 1 - 100% LEL range Resolution 1%LEL Measurement ±5%LEL error Response time = 3 s, t = 5 s Recovery time <3 s, t...
  • Page 32: Factory Settings

    3 | Description INFICON CO: t < 15 s, t <30 s <12 s, t <36 s S: t <18 s, t <66 s Recovery time <3 s, t <5 s <12 s, t <36 s <16 s, t <39 s CO: t <15 s, t...
  • Page 33 INFICON Description | 3 Parameter Factory Setting Options Screensaver (IRwin S only) enabled enabled or disabled PPM Adjustment Factor 1.0* 1.0, 1.2, 1.4, 1.6, 1.8, 2.0 Select Gas (Ex) PPM Alarm 10 PPM* 3, 5, 10, 15, 20, 25, 50,...
  • Page 34 Logging Interval (log file 1, 2, 3, 5, 10, 30 s, 1 min period) Start Mode Default (Ex or Above Default, Last Used, Ex, Ex Ground for IRwin S)* Tox, Universal (if enabled), GC Ethane Analysis Auto-rotate Screen On, Off Reminder for Ethane...
  • Page 35: Concentration Of Calibration And Test Gases

    INFICON Description | 3 3.7 Concentration of calibration and test gases Calibration routine Default gas Adjustable range Info Ex (2.x% CH 2.2 Vol% CH 2.0 - 2.7 Vol% in Select gas close to 50% LEL 0.1 Vol% increments Above Ground 10 ppm CH 8 - 15 ppm in 0.1 ppm...
  • Page 36 3 | Description INFICON Calibration routine Default gas Adjustable range Info GC Function test mix: 1 Vol% CH Not applicable Adjusts peak timing. Option 3 200 ppm C Not applicable GC Function test mix: 1 Vol% CH Not applicable Adjusts peak timing. Only option Option 4 in US English setting.
  • Page 37: Getting Started

    INFICON Getting started | 4 4 Getting started 4.1 Charging the battery WARNING Explosion hazard The equipment for charging IRwin is not ATEX certified/ex-protected. ► Never charge IRwin (all models) in potentially explosive atmospheres. WARNING Using the wrong charger could damage the explosion protection of IRwin SXnn models.
  • Page 38 Switch off IRwin S. Connect the "Charger for IRwin S, 100-240V (580-603)" or "Car adapter for IRwin S, 12V (580-602)" to the charging inlet of IRwin S. Align the red marks of the charging socket and charging plug with each other.
  • Page 39 100-240V (580-605) Certified IRwin Charging Adapter (580-604) Table 14: Accessories for charging for IRwin SXnn and IRwin S from serial number 929000705 Charging IRwin SXnn and IRwin S from serial number 929000705 Switch off IRwin SXnn. Connect the “Certified IRwin Charging Adapter (580-604)“ to the charging inlet of IRwin SXnn.
  • Page 40: Assembling The Probe System

    WARNING Risk of explosion due to not approved probe parts. The original INFICON Hand probe (part number 580-100) is the only part of the probe system that is certified for use in potentially explosive areas. Other parts of the probe system are not allowed in classified hazardous areas (Zone 0, 1, 2, Div 1 or 2).
  • Page 41 INFICON Getting started | 4 The proper functioning of the system can only be assured with compatible probe systems of the manufacturer. Attach the probe rod to the carpet or the bell probe. Fig. 10: Assembling the probe system 1 Attach the hand probe to the probe rod.
  • Page 42 4 | Getting started INFICON Install the probe rod To improve the ergonomics of the carpet probe, you can install the "probe rod extension extension". Fig. 12: Assembling the probe system 3 Remove the male quick fitting (1) from the carpet using as 17 mm wrench.
  • Page 43: Operation

    Explosion hazard IRwin S and any kind of equipment for charging (for all models), are not ex-protected. ► Do not use IRwin S or any kind of charger for S and SXnn versions in explosive atmospheres. The start-up sequence can differ slightly between the individual models.
  • Page 44: Quick Set-Up

    You can clear the entire Quick Set-up by pressing the two left buttons while the INFICON screen is displayed directly after power on. The Startup Settings screen will then appear again. Current language region setting is displayed in the About screen.
  • Page 45: Disabling Audio Alarms In Non-Ex Modes

    INFICON Operation | 5 5.2.1 Disabling audio alarms in non-Ex modes The audio alarms in all modes except Ex and ExTox can be disabled. This can be handy if you work on a site where emission is the normal condition. Advanced user login required.
  • Page 46: Activating Audio Locating Signal

    5 | Operation INFICON 5.2.5 Activating Audio Locating Signal IRwin can do set to give an audio signal to facilitate locating leaks. This particularly handy when using the Hand Probe to locate leaks on exposed pipes or equipment. The pitch / frequency of the signal increase as the gas concentration increases. The audio signal has an autorange function meaning it will drop and start to rise again when passing 100 ppm, 1000 ppm, 1% and 2.7%.
  • Page 47: Ppm Adjustment Factor

    INFICON Operation | 5 Enter the new password and press check mark. Leave the view via Reset Password Follow steps 1 and 2 in instruction for setting password and select reset. The current passwords are displayed at the bottom of the screen if you are logged in on level Advanced.
  • Page 48: Set Limits And Alarm Levels

    5.2.11 Set limits and alarm levels WARNING Explosion hazard Irwin S is not certified for use in explosive atmospheres and is also not suitable for estimating fire potential (LEL level). Alarm indications: Gas alarms are usually separated into three levels of severity:...
  • Page 49: Select Available Modes

    INFICON Operation | 5 Setting alarm ü Login as Intermediate or Advanced thresholds > > Thresholds Select desired tab: ð CxHy (gas survey and LEL alarms) ð Safety (toxic and oxygen alarms) ð xTEL (toxic exposure limits) Set the desired thresholds.
  • Page 50: Universal Fnc Test Conc / Abg Fnc Test Conc

    5 | Operation INFICON Selection of gas concentrations to use is made by changing the Setting of calibration gas concentration. See "Calibration (adjustment) [} 81]" and "Setting of calibration gas concentrations [} 85]". 5.2.14 Universal Fnc Test conc / AbG Fnc Test conc The above ground function test concentration is selected by the Quick setup if performed.
  • Page 51: Enable Dead Band Suppression (Irwin Sxnn Only)

    > Function Test > Interval (Tab) Select the function test and the time interval. ð IRwin S: You can set a time interval. ð IRwin SXnn: You can set the time interval individually for the modes “Above Ground“ / ”Universal (low range)", “Bar Hole” / ”Universal (high range)”, “Ex“, “Tox Mix“...
  • Page 52: Transfer Data To Computer

    – Survey reports with map showing position and magnitude of recorded signals. Can be commented. – GC analysis reports – Function Test reports • Live stream IRwin values to computer. IRwin Viewer is available for download on IRwin product page on inficon.com. 52 / 116 IRwin®-Operating-instructions-mina66en1-21-(2403)
  • Page 53: Paring Irwin With Computer Via Bluetooth

    INFICON Operation | 5 5.2.21.3 Paring IRwin with computer via Bluetooth ü Start IRwin and let it pass the self-test. ü Check that the Bluetooth symbol appears in the status field on IRwin display. ► Bluetooth activated Type “Bluetooth” in serch field on your Windows computer and select Bluetooth &...
  • Page 54 5 | Operation INFICON WARNING Do not use Above Ground, Universal or Universal+ modes for safety check. The accuracy is not certified for LEL measurement. ► Observe all safety precautions when alarms are given. ► Never open or enter a confined space without first testing LEL level in the Ex or ExTox mode.
  • Page 55 INFICON Operation | 5 (10 PPM) Above ground Fig. 13: “Above Ground” display Navigation buttons Navigation and on / off button Battery status indicator Mode icon Measured value, target gas and Analog bar indicating range (grey), unit measurement value (blue) and...
  • Page 56: Operating Modes Universal And Universal

    5 | Operation INFICON 5.3.3 Operating modes Universal and Universal+ These modes combine highest sensitivity for PPM levels of Methane all the way up to 100 Vol% and can be used for all applications except safety checks. The Universal+ mode is further adopted to the UK EM72 operating procedures.
  • Page 57 INFICON Operation | 5 (10 ppm) Universal Fig. 14: "Universal" display Measurement value Measured gas Unit of measured value PPM threshold Zeroing button, if needed. Zeroing Data logging activation. symbol disappears if gas is detected. Measurement range Measurement range 1 ppm - 100 Vol% methane IRwin®-Operating-instructions-mina66en1-21-(2403)
  • Page 58 5 | Operation INFICON Universal+ The Universal+ mode can be activated as followed > > General > Adv 2 (Tab) Select Universal Mode and press check mark Scroll to Universal+ and press check mark Exit by pressing  several times Fig. 15: Universal+ Display...
  • Page 59: Operating Mode "Bar Hole

    INFICON Operation | 5 5.3.4 Operating mode “Bar Hole” This mode is intended for pin-pointing leaks by measuring the concentration in the ground. The detector is ready to use when the measurement value is black and status LED is blue. No gas alarms are given in this mode.
  • Page 60: Automatic "Bar Hole" Testing

    5 | Operation INFICON Measurement range Measurement range 0.1 Vol% - 100 Vol% CH 0.1 Vol% - 20 Vol% CO You can readjust the zero point of the CH Vol% value if needed. Press lower left button if "0" appears in blue.
  • Page 61: Setting Parameters For "Auto Bar Hole

    INFICON Operation | 5 5.3.4.2 Setting parameters for “Auto Bar Hole” > > Auto Bar Hole > Settings (Tab) Measuring Duration. This is the time the detector tests the gas in each measurement. Select 10 - 20 s (5 s increments) Evacuation Duration.
  • Page 62: Operating Mode "Gc Ethane Analysis

    5 | Operation INFICON 5.3.5 Operating mode "GC Ethane Analysis" This mode is intended for determining if a gas sample is natural gas (NG) or liquefied petroleum gas (LPG), by identifying the methane, ethane and propane content. Natural gas (NG) typically contain at least 0.5 vol% of ethane in addition to a high concentration of methane.
  • Page 63 INFICON Operation | 5 IR Ethane Analysis Fig. 17: "GC Ethane Analysis" display Analog bar indicating Battery status indicator concentration range (grey) needed to perform GC Ethane Analysis and measurement value (blue) in quasi logarithmic scale Mode icon Measured value, target gas and...
  • Page 64 5 | Operation INFICON • Make a GC Ethane Analysis on the gas to be identified and compare the resulting curve to that obtained from the natural gas sample. Three consecutive tests with same result should serve as a clear result.
  • Page 65 INFICON Operation | 5 ð Purging will restart automatically if shutting IRwin down or attempting to make a new GC Ethane Analysis. Irregular GC Ethane Analysis results can in same cases occure due to the flammable gas sensor being overloaded. This could for example, happen if the detector is in high gas background during analysis or purge.
  • Page 66: Operating Mode "Plants

    5 | Operation INFICON 5.3.6 Operating mode “Plants” WARNING The accuracy of the methane measurement in the operating modes "Plants" and "House" is not certified for the LEL measurement. Observe all safety precautions when alarms are given. Never open or enter a room without first checking the LEL level in Ex or Ex Tox operation mode.
  • Page 67 INFICON Operation | 5 This mode is used for checking the gas concentration in confined spaces such as man holes, duct openings, etc. The leak detector is ready to use when measurement value is black and the status LED is blue.
  • Page 68: Operating Mode "House

    5 | Operation INFICON 5.3.7 Operating mode "House" WARNING Gas alarms are not safe in this mode! The background concentration around the detector is zeroed out. Signal will only be displayed if probe detects higher gas than the reference inlet.
  • Page 69: Operating Mode "Gas Purity

    INFICON Operation | 5 5.3.8 Operating mode "Gas Purity” This mode is intended for checking that the gas has reached a certain point in the network when a previously purged section is refilled. This mode is not available if Universal mode is activated.
  • Page 70 5 | Operation INFICON Measurement range Measurement range 0.1 - 100 Vol% methane You can readjust the zero point of the CH Vol% value if needed. Press lower left button if "0" appears in blue. 70 / 116 IRwin®-Operating-instructions-mina66en1-21-(2403)
  • Page 71: Operating Mode "Ex

    INFICON Operation | 5 5.3.9 Operating mode "Ex" This operating mode is used to check or monitor the fire or explosion risk in the working environment. The concentration of the gas is given in % LEL. A signal of 100% LEL or higher (> % LEL) means there is a risk of fire.
  • Page 72 5 | Operation INFICON ±Measurement range Measurement range 1 - 100 %LEL methane You can readjust the zero point of the %LEL value if needed. Press lower left button if "0" appears in blue. If required, change the primary target gas from methane (CH...
  • Page 73 INFICON Operation | 5 The acoustic signal and the status LED are the primary alarm signals. All gas alarms have a latching function. After a gas alarm is triggered, a check mark key appears in the upper right corner of the display.
  • Page 74: Operating Mode "Ex Tox

    5 | Operation INFICON 5.3.10 Operating mode "Ex Tox” This operation mode is used to monitor and warn for explosive and toxic gas concentrations as well as oxygen deficiency in the working area. %LEL 21.6% O 0.05% CO 0.0 ppm H S...
  • Page 75 INFICON Operation | 5 Measurement range Measurement range or C 1 - 100% LEL 0.1 - 5 Vol% 1 - 500 ppm 1 - 400 ppm 0.1 - 25 Vol% Table 18: Measurement range You can readjust the zero point of the %LEL value if needed. Press lower left button if "0"...
  • Page 76: Change Gas Type For %Lel Measurement

    5 | Operation INFICON After the leak detector has been exposed to high methane concentrations, it often displays increased measurement values in the ppm range for a few minutes. This is a normal behavior of the leak detector due to gas residues inside the leak detector.
  • Page 77: Ir Ethane Analysis And Compensation

    The resolution is 0.1% and the accuracy better than 1 Vol%. INFICON single-cell IRwin methane leak detector employs the unique INFICON single-cell wide range wide range technology to detect all concentrations from 1 ppm to 100 Vol% methane. This technology (patent- technology is very sensitive to the ethane content in the natural gas.
  • Page 78: Entering Ethane Concentration Manually

    5 | Operation INFICON 5.3.11.1 Entering ethane concentration manually If you know the current ethane concentration in your natural gas you can enter this value manually. We recommend, however, that you use the built-in IR Ethane Analysis to determine the ethane concentration as this typically varies from day to day and from location to location.
  • Page 79: Performing Ir Ethane Analysis After Automatic Reminder

    INFICON Operation | 5 5.3.11.3 Performing IR Ethane Analysis after automatic reminder You need a methane concentration in the range of 5 to 15 Vol% to perform the IR Ethane Analysis. The compensation is only needed for high gas concentrations and is therefore only needed when you have found a significant leak.
  • Page 80: Starting Ir Ethane Analysis Manually

    5 | Operation INFICON 5.3.11.4 Starting IR Ethane Analysis manually > > IR Ethane Analysis > Start (Tab) Press button next to check mark to start. Press the bell firmly down onto the place where you found the gas. ð A number between 0 and 100 appears on the display.
  • Page 81: Calibration (Adjustment)

    INFICON Operation | 5 5.3.12 Calibration (adjustment) Calibration is interpreted differently in different countries and applications. Throughout this document we therefore use the following definitions: Function Test: Exposing to given gas(es) checking if accuracy of a certain measurement is within specified accuracy and to judge weather calibration adjustment is needed.
  • Page 82 5 | Operation INFICON • Universal (low range) - weekly For calibration adjustment of Universal (low range) you need one, two or three gases as specified in section “Concentration of calibration and test gases”, see Concentration of calibration and test gases [} 35].
  • Page 83 INFICON Operation | 5 The methane calibration is possible with gas mixtures in synthetic air only. The natural background of CO (typically 400 ppm) in ambient air will offset the calibration. Fresh air exposure should, however, always be by ambient air.
  • Page 84: Standard Calibration Procedures

    5 | Operation INFICON 5.3.12.1 Standard calibration procedures Prepare calibration For a complete calibration you need the gases listed, see "Concentration of calibration and test gases [} 35]". It is not necessary that you calibrate all modes at the same time.
  • Page 85: Setting Of Calibration Gas Concentrations

    INFICON Operation | 5 • The gas flow delivered by the calibration station must be higher than the flow aspired by the leak detector. Set the gas flow to at least 80 l/h. • The gas must be delivered at ambient pressure (i.e., do not push gas into the leak detector).
  • Page 86: Adjust Calibration Reminder

    5 | Operation INFICON Set the exact gas concentration value via drop-down menus. The leak detector shows the gas to adjust in the window and the currently stored concentration appears just below the setting line. You cannot set the concentration outside the range.
  • Page 87: Performing Function Tests

    INFICON Operation | 5 5.4 Performing function tests Correct readings only if settings are correct. If the leak detector does not work correctly, it may not show every leak. ► Test the leak detector regularly as recommended. Function tests are available for all tasks of the detector.
  • Page 88 5 | Operation INFICON ü To perform the Function test, you need fresh air and either 10 ppm CH or 50% LEL in synthetic air. For IRwin SXT and SXGT you will also need the ToxMix gas mixture. Check the leak detector for visible damage.
  • Page 89: Measuring

    INFICON Operation | 5 5.5 Measuring ü Connect the leak detector to the desired probe. See "Function [} 13]". Switch on the leak detector by pressing the I/O key. Complete the self test. Make sure that the desired basic settings are correct. See "Quick Set-up [} 44]".
  • Page 90: Logging Data To File

    5 | Operation INFICON 5.8 Logging data to file IRwin can log data in all operating modes. This is especially practical to save data from leak surveys in Above Ground mode. ü You have configured the data logging, see "Configure data logging [} 50]".
  • Page 91 Red LED / Constant HW Watchdog triggered Perform hard switch off. beep Restart detector. If problems remains: Contact INFICON service. Reference Inlet Blocked Reference inlet blocked Check that nothing is blocking the inlet. Reference inlet filter Contact INFICON service. clogged...
  • Page 92 INFICON Message Possible error sources Remedy sensor error System error Restart detector. If problems remains: Contact INFICON service. S STEL level Short Term Exposure Normal function. reached, (STEL and alarm for H If warning reapperas without any gas present: limit shown,Confirm) Recalibrate TOX mode.
  • Page 93 Switch off detector and charge battery (count down to power off) voltage error System error Restart detector. If problems remains: Contact INFICON service. Detector shuts down System error Restart detector. If problems remains: Contact INFICON service. Red LED and constant HW watchdog trigged Perform hardware reset.
  • Page 94: Viewing The Service Screen

    Buzzer "hole" blocked or Dry and clean buzzer hole. in function test result filled with water Restart detector. screen. If problems remains: Contact INFICON service. "LED" has red cross in LED broken Contact INFICON service. function test result screen. RAM test error (Power Memory damaged or test Restart detector.
  • Page 95 If you need to do a hardware switch off you should restart the leak detector immediately, If the leak detector malfunctions again you will need to contact INFICON service. Press the "On / Off” key for 9 seconds.
  • Page 96: Maintenance

    6 | Maintenance INFICON 6 Maintenance The gas inlet of the instrument, as well as the probe system include various filters. These must be replaced at intervals specified in the maintenance plan. If the leak detector is used in dirty environments, more frequent replacements may be necessary.
  • Page 97: General Check Of Probe System

    Examine all parts of the probe system for external damages. Check that the rubber grip of the Hand Probe is not broken or missing. The rubber is certified anti-static and must be replaced by original INFICON spare if damaged or missing.
  • Page 98: Maintenance Of The Leak Detector

    6 | Maintenance INFICON 6.4 Maintenance of the leak detector 6.4.1 Seasonal shut-down IRwin SXT is using a small amount of power when shut off. This allows the clock to continue running and the function of oxygen sensor to be maintained for faster warm- When you want to store the leak detectorup to 3 months charge it fully.
  • Page 99: Maintenance Of The Hand Probe

    INFICON Maintenance | 6 6.5 Maintenance of the hand probe Replace the filter in the tip of the probe according to the maintenance plan, see "Maintenance plan". Be careful when checking the cigarette filter in the inlet, make sure that it is not deformed, otherwise it will loose part of its function.
  • Page 100: Maintenance Of The Bell Probe

    6 | Maintenance INFICON 6.7 Maintenance of the bell probe Fig. 24: Flexi Bell ► Change the filter according to the maintenance plan, see "Maintenance plan [} 96]". Fig. 25: Bell Probe Bell O-Ring Paper filter Filter housing Screw off the top of the filter housing.
  • Page 101: Maintenance Of The Swan Neck

    6.8 Maintenance of the swan neck Fig. 26: Swan Neck Change the filter according to the maintenance plan, see "Maintenance plan [} 96]". 6.9 Maintenance of the bar hole probe Rubber needs to be replaced when worn. Contact INFICON. Fig. 27: Bar Hole Probe IRwin®-Operating-instructions-mina66en1-21-(2403) 101 / 116...
  • Page 102: Decommissioning The Leak Detector

    ð You will then receive a return number from us. Use the original packaging when returning. Before sending the instrument attach a copy of the completed contamination declaration. See below or on the homepage of INFICON. 102 / 116 IRwin®-Operating-instructions-mina66en1-21-(2403)
  • Page 103 INFICON Decommissioning the leak detector | 7 IRwin®-Operating-instructions-mina66en1-21-(2403) 103 / 116...
  • Page 104: Disposing Of The Equipment

    7 | Decommissioning the leak detector INFICON 7.2 Disposing of the equipment The leak detector can either be disposed of by the owner or be sent to the manufacturer. The leak detector contains materials that can be recycled. This option should be exercised to prevent waste and also to protect the environment.
  • Page 105 INFICON Decommissioning the leak detector | 7 Mounting screw Red wires Black wires ü The battery is connected to the circuit board with two red and two black wires. Disconnect one of the four wires and isolate the end with electric tape.
  • Page 106: Declarations Of Conformity

    8 | Declarations of conformity INFICON 8 Declarations of conformity 106 / 116 IRwin®-Operating-instructions-mina66en1-21-(2403)
  • Page 107 INFICON Declarations of conformity | 8 IRwin®-Operating-instructions-mina66en1-21-(2403) 107 / 116...
  • Page 108 8 | Declarations of conformity INFICON 108 / 116 IRwin®-Operating-instructions-mina66en1-21-(2403)
  • Page 109 INFICON Declarations of conformity | 8 IRwin®-Operating-instructions-mina66en1-21-(2403) 109 / 116...
  • Page 110 8 | Declarations of conformity INFICON 110 / 116 IRwin®-Operating-instructions-mina66en1-21-(2403)
  • Page 111 INFICON Declarations of conformity | 8 IRwin®-Operating-instructions-mina66en1-21-(2403) 111 / 116...
  • Page 112: Index

    Index INFICON Index Bell Probe  15 Carpet Probe  15 Adjustment Factor  47 Flexi Bell  17 Alarm Function  13 Latching Function  73, 75 Gases  13 Level  48 Hand Probe  16 alive signal  72, 75 Probes  15 Ambient conditions  22 Swan Neck  17...
  • Page 113 INFICON Index Language  45 Technical Data  21 Ambient Conditions  22 Electrical Data  21 Mechanical Data  21 Maintenance  96 Physical Data  22 Bar Hole Probe  101 Time Bell probe  100 local time  45 Hand probe  99 Type certificate  23 Swan Neck  101...
  • Page 114 Index INFICON 114 / 116 IRwin®-Operating-instructions-mina66en1-21-(2403)

Table of Contents