How to do a 1-point calibration at 500ppm with your irwin methane leak detector (universal mode) (for a factory set unit)

Advertisement

Quick Links

QUICK START GUIDE
How to do a 1-point Calibration at 500ppm with your
IRwin Methane Leak Detector (Universal Mode)
(for a factory set unit)
1
Make sure your IRwin is set in Universal Mode (see
the How to set your IRwin for Universal Mode quick
start guide)
2
Make sure your IRwin is set with the right calibration
settings (see the How to set your IRwin for one point
calibration 500ppm quick start guide)
3
Press
followed by
4
Use
to navigate to Calibration and press
5
If prompted, press
password code. If not prompted, go ahead to next
step
1 of 2
and enter the Intermediate
mimg66en1-01 (1906)