How to set your methane leak detector for 1 point calibration 500ppm (universal mode)

Advertisement

Quick Links

QUICK START GUIDE
How to Set your IRwin Methane Leak Detector for
1 Point Calibration 500ppm (Universal Mode)
(for a factory set unit)
1
Make sure your IRwin is set in Universal Mode (see
the How to set your IRwin for Universal Mode quick
start guide)
2
Press
followed by
3
Use
to navigate to General and press
4
Use
to navigate to Custom and press
5
Use
to navigate to AbG Cal Setup and press
1 of 5
mimb66en1-01 (1905)

Advertisement

loading

Summary of Contents for Inficon IRwin S

  • Page 1 QUICK START GUIDE How to Set your IRwin Methane Leak Detector for 1 Point Calibration 500ppm (Universal Mode) (for a factory set unit) Make sure your IRwin is set in Universal Mode (see the How to set your IRwin for Universal Mode quick start guide) Press followed by...
  • Page 2 How to Set your IRwin Methane Leak Detector for 1 Point Calibration 500ppm to select Generic and press Press to navigate to Calibration and press to navigate to Settings and press If prompted, enter the (min) Intermediate password code and press If not prompted, go ahead to next step 2 of 5 mimd66en1-01 (1905)
  • Page 3 How to Set your IRwin Methane Leak Detector for 1 Point Calibration 500ppm Press to get to the drop-down menu Press on the first choice of the drop-down menu Navigate to far right drop-down menu and press to select Disabled and press Press again to view drop-down menu 3 of 5...
  • Page 4 How to Set your IRwin Methane Leak Detector for 1 Point Calibration 500ppm Navigate to the second choice in the drop-down menu and press to navigate the drop-down menus. Enter: 5 and press 0 and press 0 and press Navigate to far right drop-down menu and press Select Set and press Press again to view the drop-down menu...
  • Page 5 . s w e d e n @ i n f i c o n . c o m Due to our continuing program of product improvements, specifications are subject to change without notice. All trademarks are the property of the ir respective owners. ©2019 INFICON mimd66en1-01 (1905)