Chapter 6: Making Safety Interlock Connections 
Chapter 6
Making Safety Interlock Connections 
The SPDC810N is equipped with a remote interlock connector located on the rear panel. In order to enable the 
pump  laser,  a  short  circuit  must  be  applied  across  the  terminals  of  the  Remote  Interlock  connector.  This 
connection is made available to allow the user to connect a remotely actuated switch to the connector (i.e. an 
open door indicator). 
The switch that is connected to this interlock must be normally open (N.O.), meaning that it has to be closed in 
order for the unit to be enabled. If the switch is changed to an open state, the pump laser will automatically shut 
down. If the switch returns to a closed condition, the pump laser will not re‐enable until the ENABLE switch is 
pressed. 
All units shipped from Thorlabs are configured with a shorting BNC device installed in the interlock connector. If 
you are not going to use this feature then you can leave the shorting device installed and the unit will operate 
normally, as described throughout this manual. If you wish to make use of the interlock feature, you will need to 
acquire the appropriate mating connector (BNC male) and wire it to your remote interlock switch. 
 
 