Page 1
INSTRUCTIONS FOR USE Abridged version Ambu aScope ® ™ Broncho Sampler Set Family For use by trained clinicians/physicians only. For in-hospital use. For use with Ambu displaying units. ® Ambu aScope 4 Broncho Regular Sampler Set ® ™ Ambu aScope 4 Broncho Large Sampler Set ®...
Page 3
Contents Page English (Instruction for use) ..........................4-15...
In this document endoscope refers to instructions which applies to the endoscope only and system refers to information relevant for the aScope 4 Broncho and the compatible Ambu displaying unit and accessories. Unless specified otherwise, endoscope refers to all aScope 4 Broncho variants available for sampler set.
Page 5
23. The endoscope consists of parts supplied by Ambu. These parts must only be replaced by Ambu authorised parts. Failure to comply with this may result in patient injury.
2. System description The aScope 4 Broncho can be connected to the Ambu displaying unit. For information about the Ambu displaying unit, please refer to the displaying unit Instruction for use.
– Tracheal tubes for use in anaesthetic and respiratory equipment in compliance with EN ISO 5361. The aScope 4 Broncho Sampler Set has been evaluated to be compatible with the following endotracheal tubes (ETT), double lumen tubes (DLT) and Endoscopic accessories (EA) sizes.
Page 8
2.3. aScope 4 Broncho Sampler Set parts Item Function aScope 4 Broncho Flexible endoscope intended for endoscopic procedures and examination within the airways and tracheobronchial tree. Part Function Handle Suitable for left and right hand. Control lever Moves the distal tip up or down in a single plane.
For connection to a male suction adapter Suction adapter and the suction connector. Item Function Sampler bridge Connects to the aScope 4 Broncho, and creates a closed loop for sampling and/or suctioning. Part Function Suction connector For connection of suction tubing.
Page 10
UK Responsible Person. Importer (For products imported into Great Britain only). A supplementing list of symbol explanations can be found on www.ambu.com/symbol-explanation. 4. Use of aScope 4 Broncho Sampler Set Numbers in gray circles below refer to illustrations on page 2.
Page 11
Refer to the Ambu displaying unit Instruction for use for preparation and inspection of the Ambu displaying unit 2 Inspection of the image Plug in the endoscope cable connector into the corresponding connector on the compatible displaying unit. Please ensure the colours are identical and be careful to align the arrows.
4 Broncho in accordance with Instruction for use for aScope 4 Broncho. 7e To prepare for sampling of specimen, turn the flow switch counter-clockwise towards the sample container, and align it horizontally.
Page 13
The elements of the insertion cord are radio opaque. Disconnect Disconnect the endoscope from the displaying unit 10 . The aScope 4 Broncho Sampler Set is a single use device. The aScope 4 Broncho Sampler Set is considered infected after use and must be disposed of in accordance with local guidelines for collection of infected medical devices with electronic components.
Transportation aScope 4 Broncho Regular/Large Sampler Set Temperature³ [°C, (°F)] -10 – 55 (14 – 131) Relative humidity [%] 10 – 95 Atmospheric pressure [kPa] 50 – 106 Optical system aScope 4 Broncho Regular/Large Field of View [°] Direction of view [°]...
Page 15
Re-connect the sampler bridge to aScope not be connected 4 Broncho, and press the attachment lock properly to the button to secure connection. aScope 4 Broncho. Vacuum suction Ensure that the parts are properly tubing is not connected. If suction is still absent or...
Page 16
Ambu A/S Ambu Ltd Baltorpbakken 13 First Floor, Incubator 2 2750 Ballerup Alconbury Weald Enterprise Campus Denmark Alconbury Weald T +45 72 25 20 00 Huntingdon PE28 4XA ambu.com United Kingdom www.ambu.co.uk Ambu and other trademarks are trademarks of Ambu A/S.
Need help?
Do you have a question about the aScope 4 and is the answer not in the manual?
Questions and answers