Instructions
for Maintenance
SEERS CLINNOVA
Phlebotomy Couches
02-07-046-SM | ISSUE 2

Advertisement

loading
Need help?

Need help?

Do you have a question about the CLINNOVA Phlebotomy 1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Seers Medical CLINNOVA Phlebotomy 1

  • Page 1 Instructions for Maintenance SEERS CLINNOVA Phlebotomy Couches 02-07-046-SM | ISSUE 2...
  • Page 2 WARNING To avoid injury, always read this Instruction for Maintenance and accompanied documents before using the product. Mandatory to read the Instructions for Use...
  • Page 3 This Instructions for Maintenance document applies to the following products: CLINNOVA Phlebotomy 1 : Model No’s : NV9456P, NV9466P, NV9656P, NV9666P, NV9556P, NV9566P CLINNOVA Phlebotomy 2 : Model No’s : NV9476P, NV9576P, NV9676P CLINNOVA Phlebotomy Pro : Model No’s :...
  • Page 4: General Maintenance

    1. General Maintenance CONGRATULATIONS on the purchase of your new SEERS MEDICAL product. If these maintenance instructions are adhered to, the products will provide many years of safe use. If these instructions are NOT followed, your warranty could be invalid! Always keep these Instructions for Maintenance available for reference.
  • Page 5 2. Cleaning Instructions WARNING! Disconnect the couch from the mains power supply prior to carrying out any cleaning or decontaminating. DO NOT SATURATE PRODUCT! DURING CLEANING, THIS COULD IF THE PRODUCT GETS SATURATED ACCELERATE CORROSION, REDUCING PRODUCTS LENGTH OF SERVICE. General Cleaning To get rid of any dust or marks on the surface the Faux Leather MUST be wiped DAILY with a moist soft cotton cloth.
  • Page 6 Stain Removal Blot excess liquid as quickly as possible with a clean 100% cotton white absorbent cloth or sponge. If necessary, use clean warm water. Clean the entire area where the spot occurred on the cushion, arm etc. Water may dissolve dirt particles and make cleaning easier.
  • Page 7 Frequency of routine couch cleaning will depend on exact application. 1. Ensure the couch is in a suitable area for cleaning. 2. The couch must be OFF the wheels. 3. Disconnect from mains power supply if electric model. 4. Control handsets can be wiped overusing an alcohol wipe. 5.
  • Page 8: Maintenance

    3. Maintenance WARNING! Disconnect the couch from the mains power supply prior to carrying out any maintenance. All electrical maintenance should be performed by a suitably qualified person. BLUE or BROWN wires must not be connected to any earth terminal. 1.
  • Page 9 ELECTRIC MODELS ONLY 9. On electric models, check the power cable and air pipes on hand and foot switch controls, looking for any cuts, abrasions or any other deterioration that may have occurred and replace if necessary. 10. Ensure that the mains power supply plug is attached securely to the motor.
  • Page 10 Actions to be done by qualified personnel Once a year • Carry out a full test of all electrical couch positioning functions (backrest, height, etc.) • Check operation of the backrest/footrest. Check gas struts operate smoothly and correctly throughout all ranges of movement.
  • Page 11 4. Preventative Maintenance 1- Hinge Pin: Remove the e-clip, tap them out, clean and grease them as necessary and ensure there is no sign of fatigue and refit them. 2- Lubrication of all the pivot points: It is important to keep the pivot points and moving parts lubricated to ensure smooth movement.
  • Page 12: Troubleshooting

    5. TROUBLESHOOTING Fault Possible Cause Solution Electrical functions not Mains cable has become Check beneath shroud to locate working correctly disconnected, either control box and see whether the from the wall socket or power light is illuminated and that the control box the lead is connected at both ends.
  • Page 13: Mechanical Faults

    Fault Possible Cause Solution Mechanical faults Squeaking or grinding noise A tight pivot point either Locate the noisy pivot point, then a bolt or pin slacken the securing nut by ¼ turn and lubricate until noise reduces. Wheels will not lock Couch is not on level Level couch by adjusting the floor...
  • Page 14 6. Warranty and Product Lifetime SEERS MEDICAL Limited standard terms and conditions apply to this device. A copy of this is available upon request. This contains full details of warranty terms and does not affect your statutory rights. For any further information on this product, regarding spares,...
  • Page 15: Environmental Protection

    Environmental Protection This product is marked with the crossed-out wheelie bin symbol to indicate that it is covered by EU Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE). This symbol indicates that all electrical and electronic products must be taken to a dedicated collection point at end of their working life. Contact your distributor or your local authority for advice on proper disposal.