Start Up

Initial Start Up

To start up the freezer, complete the following steps:
1. Plug the freezer into the power outlet.
2. Turn ON the power switch, located behind the front grille
door, near the upper left corner.
3. Once the freezer is powered ON, the Thermo Scientific
logo is displayed on the front screen. If this is the first time
the unit is being turned ON, an initial setup must be
completed. Press the Start Setup button to initiate the
setup.
Figure 4. Main Screen
1. The first step of the setup is to select the language.
2. This screen allows you to specify the preferred display
language.
3. Once the language is selected, press the Next button.
Figure 5. Language Selection Screen
12 |
Start Up
Figure 6. Install Instructions Screen
The Install Instructions Screen shows important information
regarding:
1. Power source and ambient temperature conditions.
2. Unit positioning, spacing, and leveling.
3. Initial freezer loading.
4. Press the Next button after reviewing.
Figure 7. Unit Clock Setup Screen
1. This screen highlights the default date and time for the
selected region.
2. Press the Edit button if change is needed. This opens the
Set Region Screen.
3. If not, press the Next button after reviewing.