BA200 BA200-M (P.N. 3056610542, Revision C2, July 2013) Cooperativa Bilanciai strives to be more environmentally focused and uses recycled materials and environmentally friendly packaging where possible. As part of this initiative we have developed a short manual that uses less paper and ink to describe the main functions of your new Cooperativa Bilanciai indicator/scale.Thank you for your support of Cooperativa Bilanciai and we hope that you enjoy your new scale.
SETTING UP AND TURNING ON THE SCALES BA200 indicator: It must be connected to a load cell platform and calibrated as necessary to match the platform and user requirements. BA200 scales: There is a pillar which must be attached to the base frame first using the 4 bolts supplied. The pillar is secured to the bracket using 2 sets of screws.
Preset Tare When the indicator or scale is at zero with no weight on the platform it is possible to enter a preset tare. First zeroing the scale, enter a value using the numeric keys. Press [Tare] to tare the indicator. The value that was entered is stored as the tare value and it is subtracted from the display, leaving a negative number on the display.
CHECK-WEIGHING Check-weighing is a procedure to display an indicator or sound an alarm when the weight on the platform meets or exceeds the values stored in the memory. The memory holds values for a high limit and a low limit. Either or both the limits can be set by the user.
“rECALL” If you wish to recall any of the pre-stored limits, press [Tare] when “rEC” displayed. The display shows . Enter the number corresponding to the “rEC X” desired memory location (0 to 9) to be recalled. will be displayed for 2 seconds indicating the values stored in the location “X”...
ANIMAL (D ) WEIGHING YNAMIC Steps: “FUnC 1”, “Func 4” Press [Func]. The first option is press the [Func] key 3 more times to display “F4 PCt” “F4 Press the [Tare] key. will be displayed. Press the [Func] key to advance to the second function, AnL”...
ACCUMULATED TOTAL The indicator can be set to accumulate when a weight is added to the platform automatically or manually by pressing [Print]. See menu structure section. The accumulation function is available only during weighing. If at any time the weighing units are changed, the accumulated data will be lost. Manual Accumulation When the scale is set to manual accumulation, the weight displayed will be stored in the memory when the [Print] key is pressed and the weight is stable.
RS-232 SPECIFICATION he BA200 are both supplied with bi-directional RS-232 interface as standard. The indicator when connected to a printer or computer outputs the weight with the selected weighing unit through the RS-232 interface. Default Specifications: RS-232 output of weighing data ASCII code 9600 Baud (user selectable) 8 data bits...
CALIBRATION The scale can be calibrated using kilogram weights or using pounds weights,(for OIML models, pound is disabled) depending on the weighing unit selected at the time of calibration. Steps: To start the calibration, turn the scale off and switch on again and then press [Tare] during the self-test. Enter code number 0000 and press [Tare].
Page 17
BA200-M Models Model # BA200- 60M BA200- 150M BA200- 300M Maximum Capacity 60kg 150kg 300kg Readability 0.02kg 0.05kg 0.1kg Units of Measure Kilograms only Stabilization Time 2-3 Secs -10°C to +40°C / +14°F to +104°F Operating Temperature 230VAC 50/60Hz. Power Supply External Calibration User Selectable...
Page 18
BA200-M Models BA200 6M BA200 15M BA200 30M BA200 60M BA200 150M Model # Maximum Capacity 15kg 30kg 60kg 150kg Readability 0.002kg 0.005kg 0.01kg 0.02kg 0.05kg Units of Measure Kilograms only Stabilization Time 2-3 Secs -10°C to +40°C / +14°F to +104°F Operating Temperature 230VAC 50/60Hz.
ERROR MESSAGES During the initial power-on testing or during operation, the indicator may show an error message. The meaning of the error messages is described below. If an error message is shown, repeat the step that caused the message. If the error message is still shown then contact your dealer for support.
MENU STRUCTURE PARAMETER LAYOUT for BA200 SCALES Key functions while in this section [Tare] enter a parameter or accept the changes Press the [Func] key to [Func] move to next parameter or option enter Functions mode. [Zero] return to previous parameter or return to weighing FUNC 1 FUNC 2 FUNC 4...
WARRANTY INFORMATION SOCIETÀ COOPERATIVA BILANCIAI CAMPOGALLIANO Via S. Ferrari, 16 41011 Campogalliano (MO) - Italia Declare under our responsibility that the product: Electronic weighing terminal BA200 and its options described in this declaration complies with: Directive EMC 2004/108/CEE Directive Low voltage 2006/95/CE Conformity is censure by complying with EN 61000-6-3, EN 61000-6-4 EN 60950...
Page 22
This symbol on the weighing instrument you have purchased means that: • this electric or electronic appliance cannot be disposed of as solid municipal waste; • it must be collected separately; • it can be returned to the dealer when a new one is purchased; •...