Emerson Rosemount 3051C Quick Start Manual page 17

Pressure transmitter, flow meter
Hide thumbs Also See for Rosemount 3051C:
Table of Contents

Advertisement

February 2024
To wire the transmitter:
Procedure
1. Remove the housing cover on the field terminals side.
2. Connect the positive lead to the "
the negative lead to the "
3. Ensure full contact with terminal block screw and washer.
When using a direct wiring method, wrap wire clockwise to
ensure it is in place when tightening the terminal block screw.
NOTICE
Emerson does not recommend using a pin or a ferrule wire, as
the connection may be more susceptible to loosening over
time or under vibration.
4. Ensure proper grounding.
NOTICE
It is important that the instrument cable shield:
Do not connect the powered signal wiring to the test
terminals. Power could damage the test diode in the test
connection. Use shielded twisted pair cable for best results.
Use 24 AWG or larger wire and do not exceed 5,000 ft.
(1500 m).
5. Plug and seal unused conduit connections.
6. If applicable, install wiring with a drip loop. Arrange the drip
loop so the bottom is lower than the conduit connections and
the transmitter housing.
7. Replace the housing cover.
Quick Start Guide
Be trimmed close and insulated from touching the
transmitter housing
Be connected to the next shield if cable is routed through a
junction box
Be connected to a good earth ground at the power supply
end
" terminal (
+
" terminal.
Quick Start Guide
) and
PWR/COMM
17