Operating
6.2 Operation with potentiometer
6.2.3
Advanced calibration
Switchpoint setting - demanding applications
Typical demanding application
•
•
•
•
1. Ensure material level is well above the probe.
2. Ensure material level is well below the probe.
It is important that as much material buildup as possible is retaining on the probe.
32
Material with heavy buildup (non conductive):
High viscosity liquids
Hygroscopic/wet solids
Heavy buildup (conductive)
On site calibration
Probe covered and then uncovered, retaining
max. possible material buildup
Consult manufacturer
SITRANS LCS100 I/O Link Version
Operating Instructions, 05/2023, A5E52382716-AA