General
Power supply class III
USA & Canada: Class 2 supply necessary,
max. voltage 30 V DC
Current consumption, typ.
Electrical connection
Display
Humidity range
Temperature range, operation and storage
Enclosure
Electromagnetic compatibility
Shock and vibration
Conformity
1)
Configuration
1) With digital versions only.
Accuracy of E+E Humidity and Temperature Sensors
The measurement accuracy depends both on the performance of the measuring instrument and on the correct
installation in the application.
For best accuracy, every E+E T sensor is multipoint factory adjusted and calibrated in a highly stable RH / T
reactor. Using a high-precision dew point mirror as reference, the overall uncertainty of the factory calibration U
is minimal.
The total measurement uncertainty U
Accreditation, Evaluation of the Measurement Uncertainty in Calibration) and with GUM (Guide to the Expression
of Uncertainty in Measurement) as follows:
� �
= k ⦁
2
U
u
cal
accuracy
U
+
total
2
√3
U
.....................total accuracy incl. factory calibration
total
U
.......................the uncertainty of the factory calibration
cal
u
.................the accuracy of the measurement device
accuracy
k ...........................enhancement factor k=2, corresponding to a confidence level of 95 %.
For external calibrations, U
due to long-term drift or chemical exposure.
As designated laboratory (NMI) responsible for maintaining the National Standard for humidity and temperature in
Austria, E+E Elektronik represents the highest instance in humidity and temperature calibration.
0 - 10 V, RS485
15 - 35 V DC or 24 V AC ±20 %
4 - 20 mA (2-wire)
10 + 0.02 x R
@ 24 V DC
0 - 10 V
6 mA
4 - 20 mA
Acc. to output current
RS485
5 mA
Push-in spring terminals max. 1.5 mm
1.8" LCD, dot-matrix, 1 line, visible area 38 x 31 mm (1.5" x 1.2")
Operation
0...100 %RH non-condensing
Storage
0...95 %RH non-condensing
without display
-30...+60 °C (-22...+140 °F)
with display
-20...+60 °C (-4...+140 °F)
Material
PC (Polycarbonate), RAL 9003 (signal white), UL94 HB approved
Protection rating
IP30
EN 61326-1
FCC Part15 Class B
Tested according to EN 60068-2-64 and EN 60068-2-27
PCS10 Product Configuration Software
adapter
for E+E sensors is calculated in accordance with EA-4/02 (European
total
2
is to be used as the evaluation criterion. The calculation does not include effects
total
< V+ < 35 V DC (RL < 500 Ω)
L
@ 24 V AC
14 mA
rms
12 mA
rms
2
(AWG 16)
EN 61326-2-3
ICES-003 Class B
(free
download) and optional configuration
User Manual TES201
R
= load resistance
L
Industrial environment
Room Sensor for Temperature | 20
cal

Advertisement

Table of Contents
loading

Table of Contents