Device Status Indication and Error Monitoring

ModBus
The TES201 features a status register which contains all status and error information. It is possible to read it from
Modbus register 602 (0x259).
If a critical error occurs, all Modbus values are set to NaN (according to IEEE754 for data type FLOAT32) or to
0x8000 (INT16). Errors are displayed in bit-coded form. If an event is present, the corresponding bit is set to 1.
Measured values outside the measuring range are limited with the corresponding limit value.
Error bits
Bit 0
Bit 1
Bit 2
Bit 3
Bit 4
Bit 5
Bit 6
Bit 7
Bit 8
Bit 9
Bit10
Bit 11
Tab. 6
Device status indication register
PCS10
In the Information section, the PCS10 Product Configuration Software displays possible errors and states. Please
refer to chapter 6.2.2. Error Messages in PCS10.
BACnet
In the event of a serious error, the "Present Value" of each Analogue Input Object is set to NaN and "Reliability" is
set to NO_SENSOR. The „Event State" Property is set to FAULT.
Please refer to the TES201 PICS (Product Implementation Conformance Statement) at www.epluse.com/tes201.
Description
Temperature below lower limit
Temperature above upper limit
Temperature value = infinite
Sensing element not accessible
Display communication broken
Recommended action
1. Check the actual temperature with the
temperature output scaling of the ordered
sensor
2. Service recommended, check the repair
possibilities with the E+E service
1. Check the actual temperature with the
temperature output scaling of the ordered
sensor
2. Service recommended, check the repair
possibilities with the E+E service
1. Check the wiring
2. Service recommended, check the repair
possibilities with the E+E service
Service recommended, check the repair
possibilities with the E+E service
Service recommended, check the repair
possibilities with the E+E service