Product Description ............................6 General ..............................6 Dimensions ..............................6 Electrical Connection ..........................7 Setup and Adjustment ............................8 Product Configuration Adapter and Software for EE160 with legacy Sensing Element ....... 8 4.1.1. EE160-MxTxAx (analogue output) ........................9 4.1.2. EE160-M1TxJ3 (digital interface)........................10 Product Configuration Adapter and Software for EE160 with new Sensing Element .........11...
The user manual may not be used for the purposes of competition without the written consent of E+E Elektronik Ges.m.b.H. and may not be forwarded to third parties. Copies may be made for internal purposes. All information, technical data and diagrams included in these instructions are based on the information available at the time of writing.
For accurate measurement it is essential that the temperature of the sensing probe and mainly of the sensing head is same as the temperature of the air to measure. Avoid mounting the EE160 sensor in a way which creates temperature gradients along the probe.
Environmental Aspects PLEASE NOTE Products from E+E Elektronik Ges.m.b.H. are developed and manufactured in compliance with all relevant environmental protection requirements. Please observe local regulations for the disposal of the device. For disposal, the individual components of the device must be separated according to local recycling regulations.
User Manual EE160 3 Product Description General The EE160 sensor is optimized for cost effective, accurate measurement of relative humidity (RH) and temperature (T) in building automation. The measured data is available on ▪ Two voltage or current (2-wire) outputs ▪...
User Manual EE160 Electrical Connection Output: 0 - 10 V WARNING Incorrect installation, wiring or power supply may cause overheating and therefore personal injuries or damage to property. For correct cabling of the device, always observe the presented wiring diagram for the product version used.
4 Setup and Adjustment The EE160 is ready to use and does not require any configuration by the user. The factory setup of EE160 corresponds to the type number ordered. Please refer to the data sheet at www.epluse.com/ee160. The user can change the factory setup with the help of the Product Configuration Software and the USB configuration adapter.
Digital RS485 Modbus/BACnet Fig. 3 EE160 legacy sensing element 4.1.1. EE160-MxTxAx (analogue output) NOTICE EE160 with analogue output must be powered externally when using the EE-PCA kit. HA011059 EE160 EE-PCA Fig. 4 EE160 configuration analogue output using EE-PCA kit Fig.
4.1.2. EE160-M1TxJ3 (digital interface) NOTICE ▪ For the EE160 with digital interface, either the supply of the configuration adapter HA011066 (for configuration purposes) or an external supply (for configuration, calibration and adjustment purposes) must be used. ▪ It is not permitted to use both (external supply and configuration adapter supply) at the same time.
Fig. 9 EE160 with new sensing element If the EE160 is equipped with a new sensing element according to the picture above it must be only be conifgured or adjusted using the PCS10 Product Configuration Software (free download from www.epluse.com/pcs10...
EE160-MxTxA6 (analogue output, current version) NOTICE ▪ For the EE160 with current output an external supply must be used for configuration, calibration and adjustment purposes. ▪ It is not permitted to use the configuration adapter supply (only for voltage output).
4.2.3. EE160-M1TxJ3 (digital interface) NOTICE ▪ For the EE160 with digital interface, either the supply of the configuration adapter HA011066 (for configuration purposes) or an external supply (for configuration, calibration and adjustment purposes) must be used. ▪ It is not permitted to use both (external supply and configuration adapter supply) at the same time.
Download the PCS10 Product Configuration Software from www.epluse.com/pcs10 and install it on the PC. Connect the EE160 to the PC using the Modbus configuration adapter. Start the PCS10 software. Follow the instructions on the PCS10 opening page for scanning the ports and identifying the connected device Click on the desired setup mode from the main PCS10 menu on the left.
Digital interface default settings The recommended settings for multiple devices in a Modbus RTU network are 9600, 8, even, 1. The EE160 represents 1 unit load on an RS485 network. Device address, baud rate, parity and stop bits can be set via: ▪...
2) Register address starts from 0. 3) The choice of measurement units (metric or non-metric) is done at the time of ordering, see the ordering guide in the EE160 data sheet. It is not possible to switch from metric to non-metric or vice versa by means of the PCS.
The filter cap shall be replaced once in a while with an E+E original one. A polluted filter cap causes longer response time of the device. ▪ If needed, the sensing head can be cleaned. For cleaning intructions refer to www.epluse.com/ee160. PLEASE NOTE While replacing the filter cap take very good care to not touch or rub the sensing element.
(comparison with a reference) or adjustment (bringing the device in line with a reference). Calibration and adjustment at E+E Elektronik Calibration and/or adjustment can be performed in the E+E Elektronik calibration laboratory. For information on the E+E capabilities in ISO or accredited calibration please see www.eplusecal.com.
EN 45545-2 (HL3) 8 Conformity Declarations of Conformity E+E Elektronik Ges.m.b.H. hereby declares that the product complies with the respective regulations listed below: European directives and standards. UK statutory instruments and designated standards. Please refer to the product page at www.epluse.com/ee160...