7.0 Fault analysis and troubleshoo• ng
For your convenience, this wheelchair is equipped with an automa• c fault warning device. Once it is out of order, the controller
panel will display the pa• ern and the horn will sound. You can find where the fault is according to the table below. If the fault
persists a• er checking, please contact your service agent.
1 Flash
2 Flashes
3 Flashes
4 Flashes
5 Flashes
7 Flashes
7 Flashes (+ fl ashing
speed indicator)
8 Flashes
9 Flashes
10 Flashes
000690871.NA Rev. A
The ba• eries need charging or there is a bad
connec• on to the ba• eries.
The le• motor (M1) or encoder has a bad
connec• on.
The le• motor (M1) wiring is faulty or a stall
condi• on has been detected.
The right motor (M2) or encoder has a bad
connec• on.
The right motor (M2) wiring is faulty or a stall
condi• on has been detected.
A joys• ck fault is indicated.
A communica• on fault is indicated.
A possible controller fault is indicated.
The parking brakes have a bad connec• on.
An excessive voltage has been applied to the
controller. This is usually caused by a poor ba• ery
connec• on.
Check the connec• ons to the ba• eries. If the
connec• ons are good, try charging the ba• eries.
Check the connec• ons to the le• motor and
encoder.
Check the connec• ons to the le• motor and
encoder.
Check the connec• ons to the right motor and
encoder.
Check the connec• ons to the right motor and
encoder.
Make sure that the joys• ck is in the center
posi• on before switching on the controller.
Make sure that the joys• ck module cable is
securely connected and not damaged.
Make sure that all connec• ons are secure.
Check the parking brake, motor and controller
connec• ons are all secure.
Check the ba• ery connec• ons are all secure.