SELECTIVITY CONTROL position
Automatically selected bandwidth
(Hz)
100
316
1000
316
Hz
31.6
100
316
100
Hz
10
31.6
100
31.6 Hz
3.16
10
31.6
10
Hz
3.16
3.16
10
3.16 Hz
3.16
3.16
3.16
EFFECTIVE AVERAGING TIME
Automatically selected averaging
selected
time (sec.)
1
0.3
0.1
0.3 sec
3
1
0.3
1
sec
10
3
1
3
sec
30
30
3
10
100
100
10
30
sec
100
100
30
100
sec
100
100
100
Switch-over frequency Hz
(x1, log scale)
200
2K
Switch-over frequencies xO, x 1 and x 10 according to range.
Table 5.4. Selected EFFECTIVE AVERAGING TIME and BANDWIDTH
with 8 & T PROGRAM selector in position "8 & T Var. 4"
The actual bandwidth and effective averaging time in use at a particular
instant are indicated by the appropriate light in the cluster surrounding the
switches. The switch-over frequencies are indicated on the scale by black
spots, and are separated by half decades for programmes 1 and 2. Six
frequency ranges are defined as follows.
P
1
2Q-
63
p2
63-
200
p3
20Q-
630
p4
630- 2000
P
5
200o- 6300
p6
630Q-20000
The ranges given are for the switch position "log x1."
For programmes 3 or 4, the switch-over frequencies are separated by a
decade.
Programme 3 combines the ranges P
2
and P
3
and alsoP
4
and P
5 .
This
gives a total of four ranges.
53