Crest factor capability:
Accuracy of RMS indication
:
Linearity of RMS indication:
Output impedance:
Load Impedance:
Averaging times:
Recorder Output DC Logarithmic:
Output voltage:
Lin/Log Conversion ratio:
Dynamic Range:
Crest factor capabi I ity:
Accuracy of RMS indication:
Linearity:
Output impedance:
Load impedance:
Averaging times:
Meter:
Full scale deflection:
Up to 5 at full scale deflection decreas-
ing to
1.4
at+
10
dB.
±
0.5
dB for crest factor up to
5.
±
0.3
dB in the range
-40
dB to
0
dB,
and
±
0.5
dB over the whole range.
~5kQ.
0.1, 0.3, 1, 3, 10, 30, 100
sec.
4.5 V
at fu II deflection ( equivalent to
31.6
V
AC).
0.9 V (DC)/dB (AC).
-50
dB to
0
dB re
4.5
V.
Up to
5
at
-10
dB decreasing to
1.4
at
OdB.
±
0.5
dB for crest factors up to
5.
±0.5 dB.
<10Q.
~5kQ.
0.1, 0.3, 1, 3, 10, 30, 100
sec.
10
V R MS at Recorder output (I inear),
31.6
V at Recorder output (logarith-
mic).