Selective Measurements - BRUEL & KJAER 2010 Instructions And Applications

Heterodyne analyzer
Hide thumbs Also See for 2010:
Table of Contents

Advertisement

2. Mount the accelerometer on the measurement object (for recommend-
ed methods, see the accelerometer instruction manual).
3. Keep the 2010 as far away as possible from the vibration environment
and any other unwanted influences.
4. With the OUTPUT SECTION ATTENUATOR in
"x
1" position,
adjust the INPUT SECTION ATTENUATOR until a suitable deflec-
tion is obtained without any indicated overload. If the deflection is
insufficient, even when the INPUT SECTION ATTENUATOR is in its
10 mV range, then adjust the OUTPUT SECTION ATTENUATOR.
If the INPUT OVER LOAD light flashes, turn the INPUT SECTION
ATTENUATOR switch to a higher range.
If the OUTPUT OVERLOAD light flashes, turn the OUTPUT SEC-
TION ATTENUATOR switch to a higher range.
5.
For further information on vibration measurement the B & K booklet
"Mechanical Vibration and Shock Measurements" is available on
request.
4.4. SELECTIVE MEASUREMENTS
For selective sound, vibration or voltage measurements between 2 Hz and
200kHz using a selected bandwidth and effective averaging time, the
procedure is as follows:
34
1. Carry out the appropriate calibration procedure as described in sec-
tion 3.2.
2. Connect the appropriate transducer and preamplifier to the 2010.
3. Set the controls as follows:
FREQUENCY RESPONSE
SELECTIVITY CONTROL
EFFECTIVE AVERAGING TIME
B & T PROGRAM
OUTPUT SECTION ATTENUATOR
BANDWIDTH COMP.
GAIN CONTROL
"Selective"
As required
As required
"Manual"
"x 1"
- if necessary
"Cal."

Advertisement

Table of Contents
loading

Table of Contents