PXIe-4466/4467 Calibration Procedure
2 years
Password
NI
Note
This is the default password for all password-protected operations.
This password is site-specific.
Task
Setup
Warm Up
Verify, Adjust, and
Reverify
Verify Only
Adjust Only
Note
Estimated test times assume the user is conducting a manual
calibration. For most procedures, automating the calibration
significantly reduces test times.
Environmental Conditions
Ambient temperature
Internal DUT temperature range
Relative humidity
4
ni.com/services
Estimated Test Time
10 minutes
15 minutes
19 minutes
8 minutes
3 minutes
Verification
23 °C ± 5 °C
Tcal ± 5 °C
Operator
Connections
36
17
2
Between 20% and 80%, noncondensing
Test Points
344
171
2
Adjustment
23 °C ± 5 °C
Tcal ± 5 °C