Device Operations; Getting Ready For The Start-Up - Honeywell Vertex VC 4 User Manual

4-points continuous monitor
Hide thumbs Also See for Vertex VC 4:
Table of Contents

Advertisement

CHAPTER
4

Device Operations

Learn what you can do with your Honeywell Vertex™ VC4 System Continuous Monitor.

Getting Ready for the Start-up

Learn the Honeywell Vertex™ VC4 system start-up sequence and how to configure the analyzer module for
specific gas locations.
Honeywell Analytics loads all software on the DAq at the factory. You need to configure each point for the
target gases at your facility.
Before you begin the start-up and configuration, gather the following information:
The location to which each point is connected
Target gas at each location
Alarm levels
Ahead of the start-up sequence, make sure that the following installation steps are completed:
Sample lines
Exhaust line
AC power connection
Sample Line Dust Filters
I/O Connection
Initial Startup
Use this section to turn on your Vertex™ VC4 System and to configure the analyzer module for specific gas
locations. There are eight parts to this startup procedure:
Power Up
Start Program
Log in the HMI program
Create a configuration profile
Replace the Chemcassette® Tape
Leak Checking Sample Lines
Adjust the Flow rate
Adjust Supply Vacuum
TM
Vertex
VC4 System
35
User Manual