Honeywell Vertex VC 4 User Manual page 146

4-points continuous monitor
Hide thumbs Also See for Vertex VC 4:
Table of Contents

Advertisement

Flow is 70 cc/min less than nominal
101
102
Remaining Chemcassette® is low
Analyzer out of monitor too long
103
Run time point disable time out
104
Sample pressures greater than ambient
105
Flow is 70 cc/min more than nominal
106
Chemcassette® expired
107
No polls from HMI for 10 seconds
108
Vertex
TM
C System
Excessive point vacuum due to clog or kniked sample line
Sample line too long or ID too small
Poor gate seal
Supply vacuum insufficient (less than 7"Hg)
Condensation
Chemcassette® thickness variation
Less than preset reminder on the Chemcassette®
Analyzer out of monitor
Out of Monitor time limit too short
Alarms were manually bypassed
Run Time Point Disable time limit too short
Point pressure above atmospheric pressure while in idle
Pressure sensor Calibration error, Defective sensor
Chemcassette® thickness variation
Defective needle valve
Chemcassette® expiration date reached
Chemcassette® installed past its expiration date
Communications to HMI PC interrupted longer than 10 seconds
Check sample line and line filter.
Clean the sample line and replace filter.
Correct sample line issue.
Ensure sample line requirements are with specifications
Check nut on the optic block
Contact Honeywell Analytics Service
Plug pneumatic connector in unused slots Exhaust tubing restricted Service or switch pumps
Check internal lines for moisture
Advance the Chemcassette®.
Adjust the needle valve to achieve 200cc/min.
Contact Honeywell Analytics Service.
Replace the Chemcassette®
Reset all alarms and faults, and then return analyzer to Monitor Mode
Change time limit in Configuration Profile
Restore point to alarm active mode
Change time limit in Configuration Profile
Review sample line location.
Confirm atmospheric conditions
Contact Honeywell Analytics Service
Advance the Chemcassette®.
Adjust the needle valve to achieve 200cc/min.
Contact Honeywell Analytics Service.
Replace needle valve
Replace Chemcassette®
Replace Chemcassette®
Check Ethernet cable at rear of analyzer
Check Ethernet hub connection and operation
Check Ethernet connection to HMI PC
Contact Honeywell Analytics Service.