Calibrate Current Loop - Honeywell Vertex VC 4 User Manual

4-points continuous monitor
Hide thumbs Also See for Vertex VC 4:
Table of Contents

Advertisement

Calibrate Current Loop

The Calibrate Current Loop function calibrates the external analog devices connected to the module by generating a
2 mA to 20 mA analog signal from each individual point. If 2 mA fault indication is disabled, then the minimum
current will change from 2 mA to 4 mA.
This function provides three signal Levels methods to calibrate connected devices.
Manually: Toggling between 2 mA, 4 mA, and 20 mA signal levels
Ramp: Automatically changing output signal levels beginning at 2 mA and going up to 20 mA
Step: Step up the current loop from 4 mA to 20 mA in 1 mA increments
1. In the left navigation panel, tap OVERVIEW
2. Tap on the I/O Settings button
3. Tape on the CALIBRATE button, the Stop monitoring screen will pop up asking a user to confirm it.
4. Select the Point to output
5. Select the Signal Levels methods
6. Select the APPLY button the enable the output.
Selecting the Signal Levels - Manually
TM
Vertex
VC4 System
.
98
User Manual

Advertisement

Table of Contents
loading

Table of Contents