Formazin Calibration Procedure - Hach 2100AN User Manual

Hide thumbs Also See for 2100AN:
Table of Contents

Advertisement

• Use only the provided silicone oil. This silicone oil has the same
refractive index as the vial glass and masks minor glass differences
and scratches.
• Store the oiling cloth in a plastic storage bag to keep the cloth clean.
• If power is lost during calibration, the new calibration data is lost and
the last calibration data is used. To exit a calibration and not save the
new values, push UNITS/Exit.
• In Calibration mode, automatic range and signal averaging on are
selected. When calibration is completed, all operational modes go
back to the last settings.
• All nephelometric (turbidity units of measure) calibrations are done at
the same time.

Formazin calibration procedure

For the best accuracy, use four matched sample cells or the same sample cell for all measurements during calibration. Refer to
on page 22.
1. Remove the filter
2. Clean and inspect
assembly. Refer to
the lens of the USEPA
Change the filter
filter assembly. Refer to
assembly
on page 50.
Clean the filter
assembly
44 English
3. Hold the tab of the
USEPA filter assembly
so that the arrows point
toward the front of the
on page 50.
instrument. Push the
filter assembly fully into
the housing.
• Ratio-on and Ratio-off calibration data is measured and recorded at
the same time.
• The 7500-NTU standard does not have to be measured during
calibration if turbidity less than 4000 NTU will be measured. Push
CAL/Zero after the 4000 NTU standard is measured to complete the
calibration procedure.
• Clean the USEPA filter assembly before doing a primary calibration,
or at least every 3 months (which is the USEPA-recommended
primary calibration interval).
4. Push CAL/Zero.
5. Rinse a clean
sample cell two times
The CAL/Zero light
with dilution water. Fill
turns on, and the mode
the sample cell to the
display shows "00" .
line (about 30 mL) with
The NTU value of the
dilution water and
dilution water used in
immediately put the cap
the previous calibration
on the sample cell. Use
is shown.
the same dilution water
that was used to
prepare the formazin
standards.
Matching sample cells
6. Clean the sample
cell with a soft, lint-free
cloth to remove water
spots and fingerprints.
Do not invert the
sample cell.