Use a cell adapter

Many different test tubes, sample cells and ampules can be used to
measure samples when a cell adapter is used. Use a cell adapter when
the test tube, sample cell or ampule is less than 25 mm. Refer to
Accessories
on page 56 for the available cell adapters.
Use only test tubes and sample cells that are free of significant
scratches. Clean and apply silicone oil to all sample cells, test tubes and
ampules used with the cell adapters. Refer to
on page 19.
Note: Performance specifications may be different than shown in
on page 5 when test tubes, sample cells or ampules less than 25 mm are used.
Use a cell adapter when:
• Only a small quantity of sample is available.
• The sample to be measured is in an ampule that cannot be opened.
9. Push ENTER to
10. When
open the flow valve and
measurements are
start the fill time
complete, push FLOW.
interval.
The FLOW light turns
To do the measurement
off.
again without the fill
time interval, push
ENTER.
To do the measurement
again with the fill time
interval, push FLOW
two times, then push
ENTER.
Clean the sample cell
Specifications
11. Push and hold the
valve-control switch to
the Momentary Open
position to drain the
flow cell.
Refer to
Table 4
for minimum sample sizes.
Table 4 Minimum sample sizes
Test tube size
12 mm
13 mm
16 mm
19 mm
Sample
2.5 mL
3.5 mL
5 mL
7 mL
English 39