Measure the Gelex secondary turbidity standards
Measure the Gelex secondary turbidity standards each time the instrument is calibrated and record the new values on the Gelex vials with a water
soluble marker.
1. Clean the Gelex
2. Apply a small bead
vials with a soft, lint-free
of silicone oil from the
cloth to remove water
top to the bottom of the
spots and fingerprints.
vial.
3. Use the oiling cloth
4. Push RANGE to
to apply the oil equally
select automatic
to the surface of the
ranging.
vial. Remove the
"AUTO" is shown below
excess oil. Make sure
the Range Mode label
that the vial is almost
on the instrument.
dry.
5. Push SIGNAL AVG
6. Push UNITS/Exit to
to turn signal averaging
select the NTU
off.
measurement mode.
The SIGNAL AVG light
turns off.