3 | Controlling Monitor Quality

3 Controlling Monitor Quality

The monitor includes the Integrated Front Sensor, which enables you to perform calibration
and Grayscale Check independently.
Note
• Calibration can be performed periodically by using the schedule function of RadiCS LE that is
attached to the monitor. For more information about the setup procedure, refer to the RadiCS LE
User's Manual (stored on the CD-ROM).
• RadiCS enables you to configure the schedule settings for Grayscale Check as well as for
calibration. For more information about the setup procedure, refer to the RadiCS User's Manual.
• If you want to perform high-level quality control in compliance with medical standards /
guidelines, use optional monitor quality control software kit "RadiCS UX2".
• To adjust measurement results of the Integrated Front Sensor to those of an EIZO external
sensor (UX2 Sensor), perform correlation between the Integrated Front Sensor and the external
sensor by using RadiCS / RadiCS LE. Periodical correlation allows you to maintain the
measurement accuracy of the Integrated Front Sensor at a level equivalent to that of the external
sensor. For details, refer to the RadiCS / RadiCS LE User's Manual.

3.1 Using the SelfCalibration Function

Execute SelfCalibration and view the latest results.

3.1.1 Executing SelfCalibration

1. Select "RadiCS SelfQC" from the Setting menu, then select
2. Select "SelfCalibration" in "RadiCS SelfQC", then select
3. Select "Execute" and then
SelfCalibration is executed.
Attention
• If the power to the PC is turned off or the input signal switches while SelfCalibration is running,
the execution will be canceled.
• After SelfCalibration is complete, perform Grayscale Check at the same room temperature and
illuminance as those in the actual working environment.
Note
• The following items can be set by using RadiCS / RadiCS LE. For more information about the
setup procedure, refer to RadiCS / RadiCS LE User's Manual.
– Calibration target
– SelfCalibration schedule
– The next execution schedule is displayed in the advanced menu if the schedule was set in
RadiCS / RadiCS LE.
• The calibration target value can be confirmed by selecting the target mode in "CAL Switch Mode"
of the Setting menu.