Eizo LMM56800 Instructions For Use Manual page 65

Large monitor manager
Hide thumbs Also See for LMM56800:
Table of Contents

Advertisement

Information and manufacturer's declaration – electromagnetic radiation
The LMM56800 Large Monitor Manager is intended for use in the electromagnetic environments
noted below.
The customer or user of the LMM56800 Large Monitor Manager has to ensure that the device is
used in such an environment.
Interference im-
IEC/EN 60601
munity test
Measurement
level
Electrostatic dis-
±6 kV contact
charge (ESD)
±8 kV air
IEC/EN
61000-4-2
Fast transient
±2 kV
electric distur-
for power lines
bance (burst)
±1 kV
IEC/EN
for input / output
61000-4-4
lines
Surge voltage
±1 kV line(s)
against line(s)
IEC/EN
61000-4-5
±2 kV line(s)
against ground
Voltage dips,
<5 % V
brief interrup-
dip in V
tions, and volt-
cycles
age fluctuations
40 % V
along power
in V
supply lines
70 % V
IEC/EN
in V
61000-4-11
<5% V
in V
onds
Magnetic fields
3 A/m
with energy tech-
nology frequen-
cies (50/60 Hz)
IEC/EN
61000-4-8
Note: V
is the alternating current voltage before application of the measurement level.
T
LMM56800
Instructions for Use, 10/2019
12.2 Information on electromagnetic compatibility (EMC)
Compliance
level
±6 kV contact
±8 kV air
±2 kV
for power lines
±1 kV
for input / output
lines
±1 kV line(s)
against line(s)
±2 kV line(s)
against ground
(>95 %
<5 % V
T
) for 0.5
dip in V
T
cycles
(60 % dip
40 % V
T
) for 5 cycles
dip in V
T
cles
(30 % dip
T
) for 25 cycles
70 % V
T
dip in V
(>95 % dip
T
cycles
) for 5 sec-
T
<5% V
T
dip in V
seconds
3 A/m
Information regarding the electromag-
netic environment
It is recommended to use the device on
wood, concrete, or ceramic floors. If the
floor is made of synthetic material, the
relative humidity should be at least 30%.
The power supply quality has to corre-
spond to that of typical industrial environ-
ments or hospitals.
The power supply quality has to corre-
spond to that of typical industrial environ-
ments or hospitals.
(>95 %
The power supply quality has to corre-
T
) for 0.5
spond to that of typical industrial environ-
T
ments or hospitals.
(60 %
If the device has to continue operation
T
) for 5 cy-
even if the power supply is interrupted, it
T
is recommended to connect the device to
an uninterruptible power supply or bat-
(30 %
T
tery.
) for 25
T
(>95 %
) for 5
T
The magnetic fields with energy technol-
ogy frequencies must be in an area that
is representative of a typical location in a
typical industrial environment or hospi-
tals.
Appendix
65