Cleaning and maintenance

8.1 Cleaning

8
Cleaning and maintenance
8.1
Cleaning
NOTICE
Device maintenance, cleaning and disinfecting
• Make sure liquids do not seep into the device. Liquids that seep into the device may re‐
sult in an electric shock or failure of the device.
• The screen is extremely sensitive to mechanical influences. Absolutely avoid scratches,
shocks, etc. for this reason.
• Clean the screen when dirty using a microfiber cloth and, if necessary, a recommended
cleaning agent. Clean the housing parts with a recommended cleaning agent.
• Use only tested disinfectants.
• If a cleaning agent is sprayed directly onto the screen surface, use a microfiber cloth to
remove drops which run down before they reach the edge of the panel.
• Remove drops of liquid from the device immediately. Contact with liquids over a longer
period can cause discoloration or allow calcium deposits to form on the surface.
Note
Monitor with protective glass
If the monitor has a protective glass the glass is subject to the same cleaning and disinfec‐
tion instructions as the screen. The recommended cleaning agents and disinfectants are
the same.
Recommended cleaning agents and disinfectants
Agent class
Alcohol
Aldehyde
Chlorine derivatives
Disinfecting agents
30
Tested cleaning and disinfec‐
tion agents:
Ethyl alcohol, 96% by vol.
Mikrozid liquid
Meliseptol Rapid
Melsitt 10% by vol.
Cidex, undiluted
Terrain 0.5% by vol.
Perform 3% by weight
Morning Mist (1:64)
Terralin Protect 2% by vol.
Mikrozid Sensitive Liquid, undi‐
luted
Microbac Tissues
Further examples
Hospiset cloth
Mikrozid liquid
Aldasan 2000
Kohsolin
Gigasept FF
Quartamon Med