Page 1
USER MANUAL ATEQ ADSE 550 Version 1.0 Reference : UM-55000A-U...
Page 3
REVISIONS ATEQ ADSE 550 MANUAL We are continually working to improve our products. Therefore, the information contained in this manual, the device and the technical data are subject to change without further notice. Date Edition/Revision Reference Updated Chapters First edition...
Page 5
ATEQ is at your disposal for any further information concerning the use of the instrument under maximum safety conditions. We would like to bring to your attention that ATEQ will not be held...
Page 7
ATEQ is attached to the COFRAC and delivers a certificate following a calibration. TRAINING COURSES In the framework of partnership with our customers, ATEQ offers two types of training in order to optimise the usage and knowledge of our instruments. They are aimed at different levels of technician: •...
Page 9
PREFACE Dear Customer, You have just purchased an ATEQ instrument, we thank you for the trust you have placed on our brand. This instrument has been designed to ensure a long and unparalleled life expectancy, and we are convinced that it will give you complete satisfaction during many long years of operation.
2. ANEMOMETER & ALTIMETER/VARIOMETER ................12 Chapter 3 USE OF THE ADSE 550 1. HOW TO USE THE ADSE 550 ......................13 1.1. General ............................13 1.2. Operation ............................ 13 1.3. Preliminary Tests ........................14 ...
Page 12
Table of contents UM-55000A-U User manual ADSE 550 Page 2 / 32...
The inner tube is parallel to the fluid flow, and is open at its end, facing the flow. The pressure inside it is therefore the total pressure (Pt), the sum of the static pressure and the dynamic pressure. Differential manometer Total Static pressure Pressure UM-55000A-U User Manual ADSE 550 Page 3 / 32...
1.1. G ENERAL The ADSE 550 pressure controller is housed in a very robust case, which can be moved directly on the aircraft, on the tarmac or in the hangar, in order to carry out "in situ" testing of equipment mounted on the aircraft.
4. To reinstall the battery drawer put him back in the unit and push until it's lock Warning: Remove the batteries when the device is in storage. UM-55000A-U User Manual ADSE 550 Page 6 / 32...
Static Pitot system of the aircraft or any system of the aircraft or any other other low-pressure applications. vacuum applications. UM-55000A-U User Manual ADSE 550 Page 7 / 32...
Chapter 2 – User Interfaces Chapter 2 USER INTERFACES 1. PRESENTATION OF ADSE 550 Front panel view of the ADSE 550 UM-55000A-U User Manual ADSE 550 Page 9 / 32...
Page 20
Pressure control tap Used to control the pressure Micrometric Balance Balance Control balance between the pitot and control tap the static Micrometric Vacuum Vaccum Control Controls the vacuum source control tap UM-55000A-U User Manual ADSE 550 Page 10 / 32...
Page 21
Indicator of the vacuum in the Vacuum Tank 100 à 0kPa) tank To be operated to fill tanks Pressure/Vacuum Pump Hand Pump when they are ON (Pressure and / or vacuum) UM-55000A-U User Manual ADSE 550 Page 11 / 32...
This bench is also used for driving and adjusting altimeters, anemometers, variometers, engine pressure rating indicators, manifold pressure indicators, and other vacuum or low pressure equipment (-60 to 100kPa). Note: Several references in brackets (ex #9) refer to the ADSE 550 user interface layout. 1.2. O PERATION ...
1.3. P RELIMINARY ESTS The user must familiarize himself with the ADSE 550 before performing tests. To avoid erroneous test results and components and instruments, particular attention should be paid to the preliminary testing procedures. Each bench is fully tested before shipment. However, to ensure the integrity of sensitive tests to be performed, Pitot and static system preliminary tests should be performed before each use.
Page 25
Chapter 3 – Use of the ADSE 550 1.3.2. Pitot preliminary test 1. Switch on the bench (#6). 2. Close the valves (#8a, #8b, #8c, #8d - clockwise). 3. Fill up the pressure and vacuum tank. 4. Open the PRESSURE CONTROL valve...
Chapter 3 – Use of the ADSE 550 1.4. P ITOT YSTEM 1.4.1. Connection to the Pitot system Using the specific pneumatic connection diagram to the aircraft (see appendix), connect the Pt port of the test bench (#3) to the Pitot system of the aircraft.
Chapter 3 – Use of the ADSE 550 1.5. S TATIC YSTEM Note: If a major leak has been detected in a previous test on the Pitot system, do not perform this test before correcting the leak. 1.5.1. Connection to the Static system Using the specific pneumatic connection diagram to the aircraft (see appendix), connect the Ps port of the test bench (#4) to the Static system of the aircraft.
Page 28
Chapter 3 – Use of the ADSE 550 1.5.3. Negative Altitude Test Warning: do not exceed -1,000 ft on the altimeter (#2) 1. Close all the valves (#8a, #8b, #8c and #8d), and use the hand pump (#13) to fill up the pressure tank to 100KPa (#11).
Chapter 3 – Use of the ADSE 550 1.6. C ) / S OMBINED ITOT PEED TATIC LTITUDE The following procedure is valid for equipment such as Mach meters, engine pressure gauges, flight analyzers, etc... Note: Perform the tests referenced in sections 1.3 Preliminary tests, 1.4 Pitot System Test, 1.5 Static System Test and 1.5.2 Static Leak Test, correct the leaks found...
Page 30
Chapter 3 – Use of the ADSE 550 9. Open the VACUUM CONTROL valve (#8d) to increase the altitude at the next test point. Once the altitude is reached, close the two valves. 10. Repeat steps 3-9 until all speed tests are completed at each desired altitude.
Page 31
Chapter 3 – Use of the ADSE 550 1.6.2. Test below the local elevation 1. Switch on the bench (#6). 2. Open the BALANCE CONTROL valve (#8c) completely. 3. Gently open the PRESSURE CONTROL valve (#8b) to reach the desired altitude test point.
Page 32
Chapter 3 – Use of the ADSE 550 UM-55000A-U User Manual ADSE 550 Page 22 / 32...
Note: If there is an airspeed decrease more than 2Kts (knots), a leak exist either in the VENT valve (#8a) or in the BALANCE CONTROL valve (#8c). 6. Correct the leak before continuing the procedure. UM-55000A-U User Manual ADSE 550 Page 23 / 32...
Page 34
Note: An increase of the Altitude more than 100ft, means that there is a leak in the VACUUM CONTROL valve (#8d). 7. Correct the leak before continuing the procedure. Note: If a leak persists, keep running test for the remaining valves until there is no leak. UM-55000A-U User Manual ADSE 550 Page 24 / 32...
M: pour les ADSE av ec option multi-v oies / f or ADSE with multiple outlet option only P: pour ADSE av ec pompes internes / f or ADSE with internal pumps only T: pour ADSE av ec écrans tactiles / f or ADSE with touch screen UM-55000A-U User Manual ADSE 550 Page 25 / 32...
Chapter 4 - Service & Device Storage 3. STORAGE For a good storage of the ADSE 550, follow the giving instructions below: a. Remove the batteries from the battery drawer and the unit when storing the bench. b. Close and lock the top of the suitcase.
IPE CARRYING BAG A cover which is placed on top of the suitcase allows the user to transport the piping (and other accessories) by means of one pocket on each side. UM-55000A-U User Manual ADSE 550 Page 27 / 32...
ATEQ Customer Service 2) Contact ATEQ Customer Service The ATEQ Company disclaims any responsibility on the calibration or adjustment of the instruments that is not performed by him. UM-55000A-U User Manual ADSE 550 Page 29 / 32...
Page 40
Chapter 6 – In case of doubt about the functioning UM-55000A-U User Manual ADSE 550 Page 30 / 32...
L: 520 mm / H: 220 mm / P: 440 mm 2 LCD graphic display 128*64 (ALT-3) Display : et 114*64 (ASI-3) Backlight Green / Yellow and adjustable contrast by user. UM-55000A-U User Manual ADSE 550 Page 31 / 32...