Technical Specifications Of Rack-Mounted Housing - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

Technical specifications
10.2 General technical specifications
10.2.2

Technical specifications of rack-mounted housing

Table 10- 2
Rack-mounted housing: General technical specifications
General
Installation position
Weight
Degree of protection
Electrical characteristics
Auxiliary power supply
Power consumption
EMC interference
immunity (electromag-
netic compatibility)
during operation:
Electrical safety
Power unit cable
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Service interface (rear)
Service interface (front)
94
Horizontal
Approx. 8.5 kg
IP20 according to EN 60529
with one and/or two
OXYMAT 7 / ULTRAMAT 7
units
with one or two CALOMAT
7 units
8, with changeover contacts, parameters can be freely assigned, e.g. for measuring
range identification; load rating: 24 V AC/DC/1.7 A (total load for all 8 relay outputs in
continuous operation max. 160 W), isolated, non-sparking
8, designed for 24 V, isolated, freely parameterizable, e.g. for autoranging
Ethernet RJ 45, 100 Mbps
Ethernet RJ 45, 100 Mbps
100 to 240 V AC (nominal range of use 85 to 264 V)
50 to 60 Hz (nominal range of use 47 to 63 Hz)
Max. 280 VA
in accordance with the standard requirements of
NAMUR NE21 (05/2006) and EN 61326-1 (2013)
in accordance with the standard requirements
NAMUR NE21 (05/2006) and EN 61000-6-2:2005
in accordance with EN 61010-1, overvoltage category II
Cable diameter: max. diam. 10 mm
Core cross-section: max. 3 x 2.1 mm
Compact Operating Instructions, 11/2019, A5E35134047-04
No
2
2
Devices in explosion-proof design

Advertisement

Chapters

Table of Contents
loading

Table of Contents