SECTION 4
IEEE-488 Reference
4.2.13
0
- Operate (Model 487)
Purpose
Use to place the voltage source of the Model 487 in operate.
Format
On
Parameters
CM Place voltage source in standby.
01 Place voltage source in operate.
Default
Description
Programming
Note
Programming
Examples
Upon power-up or after a DCL or SIX comman d is received over the bus, the voltage source
will be placed in standby.
In operate, the programmed voltage source level (V command) is available at the output of the
Model 487. In standby, the voltage source is removed from the output.
When using Interlock, the unit cannot go into operate if the interlock switch is open (test fixture
lid open).
OUTPUT 722; "VW'
OUTPUT 722; "01x"
OUTPUT 722; "00x'
! Set source to 1V.
! Place Model 487 in operate.
! Place Model 487 in standby.