Maquet Prismalix 4000 Technical Manual page 108

Table of Contents

Advertisement

| 108 |
Type:
IEC 62353
compatible
control units
Type:
Type:
Type: ....................
The protective earth resistance is to be measured between point
Protective earth resistance ≤ 300 mΩ Measured : ................ mΩ
The continuity test is then to be performed between point
Continuity test
Ensure proper calibration of your measuring equipment before EACH reading.
Prismalix |
Technical manual
| EN | 01132 | 04 |
Measuring equipment used
Voltmeter
S/N: ........................
Luxmeter
S/N: ........................
Megohmeter
S/N: ........................
S/N: .........................
Electrical safety test
4
Calibration valid until: ........................
Calibration valid until: ........................
Calibration valid until: ........................
Calibration valid until: ........................
0
and the point
3
and the point
4
4
3
3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Prismalix 6000Prismalix 8000

Table of Contents