THE CS300
ABBREVIATED OPERATOR'S GUIDE
DATASCOPE IS NOW MAQUET CARDIOVASCULAR

Advertisement

Table of Contents
loading

Summary of Contents for Maquet CS300

  • Page 1 VAULT COPY THE CS300 ™ ABBREVIATED OPERATOR’S GUIDE DATASCOPE IS NOW MAQUET CARDIOVASCULAR...
  • Page 2 VAULT COPY CS300 BEzEL AND KEyPAD The CS300 provides comprehensive HELP SCREENS with easy to follow step-by-step instructions on set-up as well as alarm and advisory conditions.
  • Page 3 Abbreviated Operator’s Guide CS300 ™ VAULT COPY QUICK REfERENCE – INITIAL SET-UP 1. Establish power and verify the MAINS ON/Off and the IABP ON /Off switches are 2. Open the Helium Tank. 3. Establish ECG and Pressure. 4. If using a Sensor IAB, ensure that the IAB sensor cable (orange) has been connected.*...
  • Page 4: Operating Modes

    | 2 | CS300 Abbreviated Operator’s Guide ™ VAULT COPY IAB Sensor Low Level Pressure Output – to Bedside Monitor This output connector is used to provide an electrical signal from the IAB Sensor module to the physiologic pressure transducer input on the patient’s bedside monitor when a Sensor IAB is connected and calibrated.
  • Page 5 Trigger Source Trigger is the signal the CS300 uses to identify the beginning of the cardiac cycle. The Trigger Source keys are only active in the SEMI AUTO and Manual Operation Modes.
  • Page 6 CAUTION: If an irregular rhythm develops while using pressure trigger, the CS300 will automatically adjust deflation timing early to avoid interfering with systolic ejection. Do not attempt to adjust the deflation control. Pressure triggering is NOT recommended for use with sustained irregular rhythms.
  • Page 7 Abbreviated Operator’s Guide CS300 ™ VAULT COPY TIMING Timing refers to the positioning of inflation and deflation points on the arterial waveform. In the STANDBy mode an inflation marker appears as a highlighted interval on the Arterial Pressure Waveform. To supplement the marker, a horizontal bar is drawn beneath the Arterial Pressure Waveform.
  • Page 8 Triggering on Ectopics: The CS300 automatically deflates on and assists the ectopic beat, if the ectopic R-wave is sensed. In SEMI AUTO Operation mode, if the ectopic beat is of small amplitude, reliable triggering can be maximized if an ECG Lead is selected which minimizes the amplitude difference between the normal QRS complex and that of the ectopic beat.
  • Page 9 Prolonged Time in Standby and Thrombus on the IAB membrane: If the CS300 is inadvertently placed in STANDBy, the time in STANDBy will appear on the left side of the monitor screen. After 10 minutes an alert message will appear. The audible alert is user selectable.
  • Page 10: High Priority Alarms

    Detailed information on all Alarm and Advisory messages is provided in the CS300 Operating Instructions. The IABP utilizes both auditory and visual alarm signals to communicate the need for attention or awareness by the pump operator. ALARM messages are displayed in the ALARM MESSAGES section of the monitor display.
  • Page 11: Operator Intervention

    Abbreviated Operator’s Guide CS300 ™ VAULT COPY OPERATOR INTERVENTION for No Trigger Reattach or reposition the electrode and check the integrity of all cable/lead connections. Switch Operation to SEMI AUTO and increase the ECG Gain setting in the PUMP OPTIONS menu.
  • Page 12 | 10 | CS300 Abbreviated Operator’s Guide ™ VAULT COPY OPERATOR INTERVENTION for No Trigger Reattach or reposition the electrode and check the integrity of all cable/lead connections. Try ECG lead selections via the ECG/AP Sources menu, or select EXT. ECG, if appropriate.
  • Page 13 ™ VAULT COPY CATHETER ALARMS The CS300 continually monitors specific parameters within the closed patient pneumatic system. If a change is detected or a specific parameter is violated, IABP Assist is suspended and the High Priority Alarm tone will sound.
  • Page 14: Medium Priority Alarms

    | 12 | CS300 Abbreviated Operator’s Guide ™ VAULT COPY MEDIUM PRIORITy ALARMS Medium Priority Alarms indicate situations in which a prompt Operator response is required. This class of alarm does not suspend pumping, but may indicate a need for corrective action.
  • Page 15: Informational Messages

    Abbreviated Operator’s Guide CS300 13 | ™ VAULT COPY OPERATOR INTERVENTION for Unable to Update Timing Check integrity of all cable connections. If transducer is in use, verify transducer was not left vented. If pulse height is unusually low or over damped and transducer is in use, flush the fluid lumen per standard hospital procedure.
  • Page 16 | 14 | CS300 Abbreviated Operator’s Guide ™ VAULT COPY System Test Ok All systems passed power-up diagnostics. System Trainer Displayed when the System Trainer is being used as ECG and Pressure source. Auto R-Wave Deflate [A /SA] The patient’s rhythm is too random to predict the occurrence of the next beat.
  • Page 17: Menu Guide

    MENU GUIDE The CS300 uses menus to provide rapid access to infrequently used functions. The menu guide, below, is also located on the top of the CS300 console. MENU NAVIGATION The first step in the menu navigation is to open the desired menu by pressing the associated key.
  • Page 18: Printer Operation

    | 16 | CS300 Abbreviated Operator’s Guide ™ VAULT COPY PRINTER OPERATION The dual trace recorder provides a hard copy record of patient waveforms. Printouts follow the format defined by the setting in the Printer Preferences submenu, located in the User Preferences menu. Dual or single trace formats may be selected.
  • Page 19: Safety Disk Leak Test

    Remove the Luer plug. If IABP therapy is being started, then proceed to initial set up or return to the Power- Up Procedure section on page 2-2 of the CS300 Operating Instructions to continue the setup of the pump. 4. If the safety disk leak test fails, inspect all pneumatic connections and press the IAB fill key for two (2) seconds to repeat the test.
  • Page 20: Battery Operation

    OPERATION fROM DC-TO-AC INVERTER The CS300 can be powered from a DC-to-AC inverter if the DC source and the inverter meet the specifications defined in the CS300 “Operating Instructions” on page 5-4. The DC source and inverter should be checked for proper operation by qualified maintenance personnel prior to emergency use.
  • Page 21 The Universal Transport System (UTS) is specifically designed for air and vehicular transport. The UTS version mechanically attaches to a docking station for high strength mounting. Refer to additional detail on the UTS on page 5-9 of the CS300 Operating Instructions.
  • Page 22 | 20 | CS300 Abbreviated Operator’s Guide ™ VAULT COPY REMOVING PUMP CONSOLE fROM THE CART The console can be removed from the cart with or without the battery pack attached. When removing the pump console or the monitor from the cart or returning them to the cart, ensure that the wheels of the cart are in the locked position.
  • Page 23 Abbreviated Operator’s Guide CS300 21 | ™ VAULT COPY To attach the monitor to the pump console: Detach the monitor from the cart handle by pressing the button on the rear of the monitor (fig. 4). Attach the monitor to the top of the pump console. Make sure the monitor is securely attached before moving the system.
  • Page 24 VAULT COPY Datascope Corp. Cardiac Assist Division 15 Law Drive Fairfield, NJ 07004 Phone: +1 973 244 6100 for local contact: Please visit our Website www.datascope.com www.maquet.com...

Table of Contents