If the password is wrong the display will return to measurement screen. The screen will display
menu options when correct password entered at last digit:
Select Calibration. Screen will show calibration options:
It would be better to first set reference gas details and calibration cycle period as these values
cannot be altered once calibration started for zero or span. Zero or Span calibration can be
performed separately at any time. It is recommended to perform Zero calibration first and
perform Span calibration following zero calibration completion.

Set Reference Gas details

During calibration technician should use a certificated specific gas. The details of the gas should
be entered before starting the calibration. Otherwise detector program will use default values
entered at factory. If the calibration gas is different than default gas specifications the
calibration will fail or will be wrong. Select fourth item 'Reference Gas' on calibration menu to
set details of the gas you will use during calibration. The default value is % 30 LEL will be
displayed:
The first digit will start blinking on screen. You can change the value of each digit with up and
down keys. Once set, confirm value with Enter key and move to the next digit.
PRS-UM-SV-EN-Rev1.0-16.12.2022
1. Alarm Settings
2. Calibrate
3. Configure
4. Information
5. Test
6. Exit
1. Set Zero
2. Set Span
3. Cancel Cal
4. Reference Gas
5. Cal. Interval
6. Previous Menu
Reference Gas
Conct.: 030 %LEL
Type: METHANE
Page 32 / 41