Thermo Scientific TSG Series Installation Manual page 26

Laboratory & pharmacy refrigerators
Table of Contents

Advertisement

Table 11. Programming Parameters
Par.
Menu
Desc.
St
Ctl
Temperature control set point
/ 5
Pro
Unit of measure: 0 = °C; 1 = °F
Hb
CnF
Enable buzzer: 0 = disabled; 1 = enabled
AHA
ALn
High temperature alarm threshold
ALA
ALn
Low temperature alarm threshold
rL2
n
Minimum temperature value in the monitored period
nt
rH2
n
Maximum temperature value in the monitored period
nt
rt
n
Monitoring period
nt
Sth
CtL
Humidity control set point
HU
CtL
Humidity level
HAn
HcP
Number of type HA alarms
HFn
HcP
Number of type HF alarms
rHP
HcP
Reset temperature alarm log
Ad
ALn
Delay time for high and low temp. alarms (AH,AL)
Add
ALn
High temperature alarm bypass time for door open
rSA
ALn
Reset alarms
PDU
CnF
User password
/cc
Pro
Control Probe calibration offset
cr
Pro
Product Probe calibration offset
Laboratory & Pharmacy Refrigerators
Def.
Min
Max
5
4
7
0
0
1
1
0
1
10
0
555/ 999
2
0
200/ 360
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
0
0
6
0
0
6
0
0
1
0
0
240
1
1
240
0
0
1
0
0
999
-
-20
20
-
-20
20
UOM
°C/°F
-
-
°C/ °F
°C/ °F
°C/ °F
-
-
-
-
-
-
-
min
min
-
-
°C/ °F
°C/ °F
Operation |
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents