Service and maintenance
7.2 Maintenance and repair work
7.2.5
Function test
Function test
A frequent function test may be required depending on the application.
Observe all relevant safety precautions related with a safe work depending on the
application (e.g. hazardous locations, hazardous material, electric safety, process pressure).
This test does not prove if the unit is sensitive enough to measure the material of the
application.
A function test is completed by touching the probe with appropriate means (e.g. grounded
metal plate or hand) and monitoring if a correct change of the signal output from
uncovered to covered occurs.
7.2.6
Production date
Production date
The production date can be traced by the serial number on the nameplate. Please contact
the manufacturer or your local distributor. For ATEX/UKEX/IEC-Ex versions, see Approvals
(Page 68).
7.2.7
Impermissible repair and maintenance of the device
Impermissible repair and maintenance of the device
• Repair and maintenance must be carried out by Siemens authorized personnel only.
7.2.8
Spare parts
Spare parts
All available spare parts are stated in the selection list.