Download Print this page

Caen 803 Series Manual page 11

803x series 8-16 ch. programmable hv power supplies

Advertisement

Remote communication control section
HV Channel Output
WARNING! These connectors produce extremely hazardous high voltages at a potentially lethal
current level; never connect or disconnect the HV OUT connector with the power ON/OFF
switch ON; always switch power OFF and wait at least 30s before connecting or disconnecting
HV cables.
AC Input (DT803x, R803x)
IEC 60 320 Socket with switch; to be connected to Mains 100 - 240 Vac (50 - 60 Hz) via provided power
cord.
Two Fuses: 6.3x32 (8 channels: 1A; 16 channels: 2A); Retarded 250VAC
11
CAEN
Electronic Instrumentation
USB
B TYPE USB
ETH
10/100 Base-T female connector
Channel Output: 8031 series: BNC RADIALL R141.557.000
Electrical: Impedance 50Ω; Operating temp. -65/+165°C;
Frequency 0-4 GHz; VSWR 1.06+0.006 xF(GHz) Max; Insertion loss
.025*√F(GHz) dB Max; RF leakage – (57 - F(GHz)) dB Max; Voltage
rating 500Veff Max ; Dielectric withstanding voltage 1500Veff
min; Insulation resistance 5000MΩ min;
Mechanical: Center contact retention: Axial force – Mating end
18 N min; Axial force – Opposite end 18N min; Torque 2.8 N.cm
min Recommended torque: Panel nut 250 N.cm; Mating life 500
Cycles min Weight 8.000 g
Channel Output: 8032-33-34-34H series: SHV RADIALL R317580;
Impedance: 50 Ohm; Frequency range: 0 –
2 GHz; VSWR: <1.20 + 0.3 F (GHz) – (plug and jack); Test voltage:
10kV DC – 1mn (unmated
connectors); Ratings: 12kV DC – 1mn (mated pairs); Current
rating: 10 A
MON
AMP 280371-2; The test points allow to monitor the
Channel Output Voltage, Current and Temperature.
V/I analog monitor output; range: 0 ÷ +5V; proportional to output
voltage and current.
T: input for PT1000 temperature probe (8031 series only).
803x Series 8-16 Ch. Programmable HV Power Supplies
USB2.0 compliant

Advertisement

loading