Page 1
Technical Information Manual Revision n.5 2 May 2005 MOD. A 1732 - A 1832 HIGH VOLTAGE BOARDS MANUAL REV.5 NPO: 00101/00:17XXx.MUTx/05...
Page 2
User. It is strongly recommended to read thoroughly the CAEN User's Manual before any kind of operation. CAEN reserves the right to change partially or entirely the contents of this Manual at any time and without giving any notice.
INTRODUCTION.................................3 CAEN U ................3 NIVERSAL ULTICHANNEL OWER UPPLY YSTEM SY 1527 ..................5 ECHNICAL PECIFICATIONS ABLE OF THE SYSTEM MOD. A 1732 - A 1832 HV BOARDS.........................6 ............................6 UNCTIONAL DESCRIPTION ........................7 HANNEL HARACTERISTICS ABLE ................................8 RONT ANEL ............................9 ECHNICAL PECIFICATIONS 2.4.1 Packaging ..............................10...
1.1 - T SY 1527 ............5 ABLE ECHNICAL SPECIFICATIONS OF THE MAINFRAME GENERAL 2.1 - C . A 1732 / 1732B - A 1832 / 1832B HV B .....7 ABLE HANNEL CHARACTERISTICS OF THE OARDS 4.1 - F ........................16 ABLE...
The SY 1527 system is the fully equipped, large scale experiment version of a new line of power supply systems which represent CAEN's latest proposal in the matter of High Voltage and Low Voltage Power Supplying. This system outlines a completely new...
Page 6
All the operational parameters are stored in a non- volatile memory (EEPROM) to be still available after Power-Off. The parameters can be controlled either via CAEN traditional built-in links (RS232, H.S. CAENET) or via CERN- approved Fieldbuses or via Ethernet (TCP/IP). Programmable handling of parameters and errors is available as well.
Document type: Title: Revision date: Revision: User's Manual (MUT) Mod. A1732-A1832 High Voltage Boards 02/05/2005 Technical Specifications Table of the SY 1527 system Table 1.1 - Technical specifications of the SY 1527 mainframe: general - 19"-wide, 8U-high Euro-mechanics rack; Packaging - Depth: 720 mm.
(A 1732BP – A 1832BP) or negative polarity (A 1732BN – A 1832BN). The Mod. A 1732 - A 1732B output channels have 6 kV / 1 mA full scale range (dip- switch selectable); the Mod. A 1832 - A 1832B output channels offer either 6 kV / 1 mA or 6 kV / 200 µA full scale range (dip-switch selectable);...
Document type: Title: Revision date: Revision: User's Manual (MUT) Mod. A1732-A1832 High Voltage Boards 02/05/2005 +5 ÷ +65 °C Threshold temperature parameters (TMIN, TMAX): Front Panel NPO: Filename: Number of pages: Page: 00101/00:17XXx.MUTx/05 A1732B-A1832B_REV5.DOC...
12 CH POSITIVE POSITIVE 6 KV 1/0.2 mA 6 KV 1/0.2 mA Fig. 2.1 – Mod. A 1732 – A 1832 (LEFT) and Mod. A 1732B – A 1832B (RIGHT) front panel Technical Specifications NPO: Filename: Number of pages: Page: 00101/00:17XXx.MUTx/05...
Mod. A1732-A1832 High Voltage Boards 02/05/2005 2.4.1 Packaging The Mod. A 1732 – A 1832 modules are housed in a 5 TE-wide, 6U-high mechanics. The Mod. A 1732B – A 1832B modules are housed in a 10 TE-wide, 6U-high mechanics. 2.4.2 External connections The location of all components of the front panel is shown in Fig.
Document type: Title: Revision date: Revision: User's Manual (MUT) Mod. A1732-A1832 High Voltage Boards 02/05/2005 Type: red LED for positive polarity version; yellow LED for negative polarity version 2.4.4 Other components VMAX trimmer Function: it allows to adjust the hardware maximum voltage VMAX common to all the channels.
Document type: Title: Revision date: Revision: User's Manual (MUT) Mod. A1732-A1832 High Voltage Boards 02/05/2005 3. Safety information and installation requirements General safety information This section contains the fundamental safety rules for the installation and operation of the boards. Read thoroughly this section before starting any procedure of installation or operation of the product.
Refer to Manual Installation The Mod. A 1732 – A 1832 are single-width boards which can be inserted in any slot of the SY 1527 crate. The Mod. A 1732B – A 1832B are double-width boards which occupy two SY 1527 slots each. At power ON the SY 1527 system processor will scan all the slots in the crate to find out where the module is plugged and what kind of module it is.
02/05/2005 4. Operating modes The Mod. A 1732 – A 1832 and A 1732B – A 1832B boards can be controlled, either locally or remotely, through the SY 1527 software interface. For details on SY 1527 system operation, please refer to the User's Manual of this product. The following sections contain a description of commands available for the board control and status monitoring.
Document type: Title: Revision date: Revision: User's Manual (MUT) Mod. A1732-A1832 High Voltage Boards 02/05/2005 • (channel in OVERVOLTAGE condition) • (channel in UNDERVOLTAGE condition) • EXTTRIP (channel OFF due to external TRIP line signal) • INTTRIP (channel OFF due to internal OVERCURRENT condition) •...
Page 18
Document type: Title: Revision date: Revision: User's Manual (MUT) Mod. A1732-A1832 High Voltage Boards 02/05/2005 T A B C Dip Switch (side view) 4 3 2 1 Out 0 Out 1 Out 2 Channels Channels Channels Out 3 Out 4 Out 5 Controller Out 6...
Page 19
Document type: Title: Revision date: Revision: User's Manual (MUT) Mod. A1732-A1832 High Voltage Boards 02/05/2005 (FSR = 6 kV / 1 m A) Dip-switch 2 = ON Dip-switch 1, 3, 4 = OFF μ (FSR = 6 kV / 200 Dip-switch 1, 2, 3, 4 = OFF FRONT PANEL...
Document type: Title: Revision date: Revision: User's Manual (MUT) Mod. A1732-A1832 High Voltage Boards 02/05/2005 APPENDIX A Notes for Mod. A1832 E Each A1832 E channel is provided with an "HV EN" input (on the right side of the SHV connector) that disables the channel when the two pins are not shorted (jumper removed).