No part of this manual may be reproduced in any form or by any means, electronic, mechanical, recording, or otherwise, without the prior written permission of CAEN SpA. CAEN will repair or replace any product within the guarantee period if the Guarantor declares that the product is defective due to workmanship or materials and has not been caused by mishandling, negligence on behalf of the User, ...
CAEN Electronic Instrumentation General description Overview Fig. 1: DT54xx USB High Voltage Power Supply The DT54xx series modules house one High Voltage Power Supply Channel (1 W max), available with either positive or negative output polarity. The units are supplied and controlled via USB link; HV output is delivered through SHV connector. The HV output RAMP‐UP and RAMP‐DOWN rates may be selected independently for each channel with 1 V/s steps. Safety features include: OVERVOLTAGE and UNDERVOLTAGE warning when the output voltage differs from the programmed value (±2%) Programmable VMAX hardware protection limit OVERCURRENT detection: if a channel tries to draw a current larger than its programmed limit, it is signalled to be in "overcurrent" and works as a current generator; output voltage varies in order to keep the output current lower than the programmed value. "Overcurrent" lasting more than set value (1 to 999 seconds) causes the channel to "trip". Output voltage will drop to zero either at the Ramp‐ down rate or at the fastest available rate, depending on Power Down setting; in both cases the ...
CAEN Electronic Instrumentation Technical specifications Packaging and Compliancy 3 The unit is a Desktop module housed in a 12.7x81x43 mm alloy box. Power requirements The module is powered by the host PC USB link. Front and Back panel Fig. 2: DT54xx front and back panel DT54xx USB High Voltage Power Supply 6 ...
CAEN Electronic Instrumentation External components HV Channel Output Fig. 3: HV Out connector HV OUT SHV (RADIALL R 317 580) HV Channel Output Earth Fig. 4: Earth connection RITEL 10mm industrial terminal 24 10 00 3 IMPORTANT: in order to reduce noise level and for safety reasons, always connect to Earth the unit through this terminal, before connecting HV Output to the load. Interlock section Fig. 5: Interlock connector AMP 280371‐2 connector & RED LED Input signal Interlock signal (channels hardware disabled) Fig. 6: INTERLOCK electrical scheme A schematic diagram of the Interlock input is shown in the figure above, where the diode is part of opto‐ coupler stage. Interlock means that channel is hardware disabled. The interlock operation is as follows: DT54xx USB High Voltage Power Supply 7 ...
CAEN Electronic Instrumentation contact open INTERLOCK voltage level (0÷1V, ~5mA current) between pin 2 and pin 3 INTERLOCK short circuit pin 1 with pin 2, and pin 3 with pin 4 ENABLED voltage level (4÷6V, ~5mA current) between pin 2 and pin 3 ENABLED The front panel Interlock LED is ON when the INTERLOCK is active; as INTERLOCK is active, channels are turned off at the fastest available rate, regardless the RAMP DOWN setting. USB Port Fig. 7: USB Optical Connector B TYPE USB port USB2.0 compliant (200cm max cable length) Status LEDs Fig. 8: LC Optical Connector ON GREEN LED lights up when the channel is active OVC RED LED lights up when channel draws a current larger than Iset (OVC detected) POL RED/YELLOW LED (always ON) lights up RED (positive channel) or YELLOW (negative channel) HV enable Fig. 9: Channel enable Lockable Toggle Switch: locks into selected position Channel Enable and Disable by pulling actuator, prevents accidental switching If “disabled”, the channel is turned off with the fastest available rate (no Ramp) HV Max Fig. 10: HVmax trimmer ...
CAEN Electronic Instrumentation Technical specifications table Table 2: DT54xx Series Channel technical specifications Model DT5470 DT5471 DT5472 3 Packaging Desktop module housed in a 12.7x81x43 mm alloy box Output channels Positive and Negative Polarity available Maximum output power 0,5W Out Range 5KV / 200µA 3KV / 500µA 500V / 1mA Vset resolution 500mV 100mV Vmon resolution 500mV 100mV Iset resolution 40 nA 100nA 200nA Low range 4nA 10nA 20nA Imon Resolution High range ...
The inspection should confirm that there is no exterior damage to the unit, such as broken knobs or connectors, and that the panels are not scratched or cracked. Keep all packing material until the inspection has been completed. If damage is detected, file a claim with carrier immediately and notify CAEN. Before installing the unit, make sure you have read thoroughly the safety rules and installation requirements, then place the package content onto your bench; you shall find the following parts: ...
The connection can be performed via terminal emulator, such as HyperTerminal; select the used USB (COM) port and configure it as follows: baud rate 9600 Data bits: 8 Parity: none stop bit: 1 Flow control: hardware In order to launch the communication, type CAEN and then <Enter>; as the communication is established, the Main Menu will be displayed: Fig. 11: Main Menu Type D to set/monitor channel parameters Type F to format the EEPROM Type U to upgrade the firmware ...
CAEN Electronic Instrumentation Firmware upgrade By typing U it is possible to access the firmware upgrade menu: Fig. 13: Firmware Upgrade Menu If <y> is typed, then the Flash is erased; At this point it is necessary to upload the updated firmware. If "HyperTerminal" is used, it is necessary to perform "Transfer" and "Send Text File" operations by selecting the new firmware file. Format EEPROM By typing F it is possible to access the format EEPROM menu: Fig. 14: Format EEPROM Menu DT54xx USB High Voltage Power Supply ...
CAEN Electronic Instrumentation USB Communication Protocol The following Protocol allows to communicate with the module. The Protocol is based on commands made of ASCII characters strings. The recognized values are as follows: Version DT5472 DT5471 DT5470 Parameter unit VSRES V 1 VMRES V 1 RAMPMAX V/s 500 RAMPMIN V/s 1 ...
CAEN Electronic Instrumentation $CMD:MON,PAR:ISET<CR,LF> Read out ISET value $CMD:MON,PAR:VMON<CR,LF> Read out VMON value $CMD:MON,PAR:IMON<CR,LF> Read out IMON value $CMD:MON,PAR:RUP<CR,LF> Read out RUP (RAMP UP) value $CMD:MON,PAR:RDW<CR,LF> Read out RDW (RAMP DOWN) value $CMD:MON,PAR:PDWN<CR,LF> Contains 'RAMP' or 'KILL' $CMD:MON,PAR:IMRANGE<CR,LF> Contains 'HIGH' or 'LOW' $CMD:MON,PAR:TRIP<CR,LF> Read out TRIP value $CMD:MON,PAR:MAXV<CR,LF> Read out Vmax hardware value Vmax can be set via hardware. $CMD:MON,PAR:IMAX<CR,LF> Read out ISET maximum value $CMD:MON,PAR:STAT<CR,LF> Read out Status word Bit 0 → ON 1 : ON 0 : OFF Bit 1 → RUP 1 : Channel Ramping UP Bit 2 → RDW...
Page 15
CAEN Electronic Instrumentation $CMD:SET,PAR:RUP,VAL:xxx.<CR,LF> Set RUP (RAMP UP) value $CMD:SET,PAR:RDW,VAL:xxx.<CR,LF> Set RDW (RAMP DOWN) value $CMD:SET,PAR:TRIP,VAL:xxx.xx<CR,LF> Set TRIP value $CMD:SET,PAR:PDWN,VAL:RAMP/KILL<CR,LF> Set power down mode 'RAMP' or 'KILL' $CMD:SET,PAR:IMRANGE,VAL:HIGH/LOW<CR,LF> 'HIGH' or 'LOW Set range $CMD:SET,PAR:ON<CR,LF> ON Set CHANNEL $CMD:SET,PAR:OFF<CR,LF> OFF Set CHANNEL $CMD:SET,PAR:BDCLR<CR,LF> 'VAL' field contains BDCLR (CLEAR) DT54xx USB High Voltage Power Supply 15 ...
CAEN Electronic Instrumentation LabVIEW Instrument Driver CAEN provides a set of VI’S developed for LabVIEW™ 2009 and later releases (LabVIEW™ is a Trademark of National Instruments Corp.) that allow to configure and monitor all parameters of DT54xx HV Power Supply modules. A LabView Demo Control Software is also provided; see p. 38. System requirements and installation setup Host PC running LabVIEW 2009 or later releases and NI‐VISA Run‐Time Engine 5.3 In order to install the DT54XX LabVIEW Instrument Driver: • Go to CAEN web site in the “Software” area of DT54xx HV Power Supply page. • Download the DT54XX LabVIEW Instrument Driver installation package • Extract files related to your host PC • Follow the Set Up wizard instructions Function Description The present sections describe in detail the DT54XX LabVIEW Instrument Driver available items; the data types are the following: Cluster (error codes) Cluster (various) Cluster (numeric) Input / Output name Boolean String 32bit integer 32bit unsigned integer numeric 16bit unsigned integer numeric ...
CAEN Electronic Instrumentation CAEN DT54XX.lvlib:Read out MAXVSET.vi This function returns MaxV value Connector Pane Controls and Indicators error in (no error): error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. Status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. Code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. Source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name BoardHandle is a reference to an open board. error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. Status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a ...
CAEN Electronic Instrumentation Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. Status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name BoardHandle is a reference to an open board. get more? : If get more? is TRUE, get more constant information releated a this parameter and return these in a cluster . The default is FALSE. If is TRUE the VI require more time to execution. error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a ...
CAEN Electronic Instrumentation status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name BoardHandle is a reference to an open board. error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error out control on the front ...
CAEN Electronic Instrumentation VISA resource name BoardHandle is a reference to an open board. CMD Command type: MON (monitor) or SET. VAL If required, value to set PAR error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error out ...
CAEN Electronic Instrumentation Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name out Module Info is a cluster with some information on the board Name Firmware Release Serial Num. ...
CAEN Electronic Instrumentation information about the error. source: source describes the origin of the error or warning. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name out VMon value (V) CAEN DT54XX.lvlib:Initialize.vi Establishes communication with the instrument. Therefore, call this VI before calling other instrument driver VIs for this instrument. Generally, you need to call the Initialize VI only once at the beginning of an application. Connector Pane Controls and Indicators error in (no error): error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front ...
CAEN Electronic Instrumentation warning or that no error occurred. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name out A copy of the reference to the instrument in use. The reference to the resource name is created in the Initialize VI. Wire the VISA resource name out terminal of the Initialize VI to subsequent VISA resource name input terminals. You chain instrument driver VIs together by wiring the VISA resource name out terminal to subsequent VISA resource name input terminals. CAEN DT54XX.lvlib:Configure VSET.vi This function allows to configure Vset. Connector Pane Controls and Indicators error in (no error): error in can accept error information wired from VIs previously called. Use ...
CAEN Electronic Instrumentation CAEN DT54XX.lvlib:Read out POWER DOWN.vi This function returns Power Down mode Connector Pane Controls and Indicators error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name BoardHandle is a reference to an open board. ...
CAEN Electronic Instrumentation CAEN DT54XX.lvlib:Read out TRIP.vi This function returns Trip time Connector Pane Controls and Indicators error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name BoardHandle is a reference to an open board. get more? : If get more? is TRUE, get more constant information releated a this parameter and ...
CAEN Electronic Instrumentation CAEN DT54XX.lvlib:Read out RAMP DOWN.vi This function returns Ramp Down rate Connector Pane Controls and Indicators error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name BoardHandle is a reference to an open board. error out: error out passes error or warning information out of a VI to be used by other VIs. ...
CAEN Electronic Instrumentation status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name BoardHandle is a reference to an open board. get more? : If get more? is TRUE, get more constant information releated a this parameter and return these in a cluster . The default is FALSE. If is TRUE the VI require more time to execution. error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a ...
CAEN Electronic Instrumentation together by wiring the VISA resource name out terminal to subsequent VISA resource name input terminals. error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a ...
CAEN Electronic Instrumentation error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out indicator on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error out indicator on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error out indicator on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source string describes the origin of the error or warning. Right‐click the error out indicator on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name out CAEN DT54XX.lvlib:Read out IMON.vi This function returns Imon value Connector Pane Controls and Indicators error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from ...
CAEN Electronic Instrumentation control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name out IMon value (uA) IMon constants Range Digits of precision CAEN DT54XX.lvlib:Configure RAMP DOWN.vi This function allows to configure Voltage Ramp Down rate. Connector Pane Controls and Indicators error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and ...
CAEN Electronic Instrumentation CAEN DT54XX.lvlib:Channel Status.vi Returns status for channels or a desired channel. Connector Pane Controls and Indicators error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name BoardHandle is a reference to an open board. error out: error out passes error or warning information out of a VI to be used by other VIs. ...
CAEN Electronic Instrumentation MAXV indicates whether the channel is in MAXV protection TRIPPED indicates whether the channel will shutdown due to the trip time VCC FAIL indicates whether there is a fault in the internal power of the channel OVT indicates whether the channel is Over Temperature CAL_CHKERR unused DISABLED indicates whether the channel is in disabled via switch INTLCK indicates whether the channel is in INTERLOCK via frontal panel CAEN DT54XX.lvlib:Read out POLARITY.vi This function returns channels polarity Connector Pane Controls and Indicators ...
CAEN Electronic Instrumentation CAEN DT54XX.lvlib:Configure IMON RANGE.vi This function allows to enable Current Monitor Zoom 10x (optional) Connector Pane Controls and Indicators VISA resource name BoardHandle is a reference to an open board. IMon range (LOW) Current Monitor Zoom 10x (optional). The admitted values are: HIGH or LOW. error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in ...
CAEN Electronic Instrumentation this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out indicator on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error out indicator on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error out indicator on the front ...
CAEN Electronic Instrumentation Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name out CAEN DT54XX.lvlib:Configure RAMP UP.vi This function allows to configure Voltage Ramp Up rate. Connector Pane Controls and Indicators error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain ...
CAEN Electronic Instrumentation CAEN DT54XX.lvlib:Configure TRIP.vi This function allows to configure Ouput Trip time. Connector Pane Controls and Indicators error in (no error) : error in can accept error information wired from VIs previously called. Use this information to decide if any functionality should be bypassed in the event of errors from other VIs. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. Trip (0.0 s) Maximum time an "overcurrent" is allowed to last expressed in seconds ...
Page 37
CAEN Electronic Instrumentation warning or that no error occurred. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. source: source describes the origin of the error or warning. Right‐click the error in control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. VISA resource name is a reference to an open board. error out: error out passes error or warning information out of a VI to be used by other VIs. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. status: status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred. Right‐click the error out control on the front panel and select Explain Error or Explain Warning from the shortcut menu for more information about the error. code: code is the error or warning code. Right‐click the error out control on the front ...
CAEN Electronic Instrumentation DT54xx LabView Control Software The present section describes an application, provided as example of DT54XX LabView Instrument Driver usage; it is provided both as executable file and source code. This demo control software allows to set and monitor, through a Graphical User Interface, all the unit’s functional parameters. Control Software System requirements and installation setup Host PC running LabVIEW 2009 or later releases and NI‐VISA Run‐Time Engine 5.3 LabVIEW™ is a Trademark of National Instruments Corp. In order to install the DT54xx LabView Control Software: Go to CAEN web site in the DT54xx page Download DT54xx Control Software package Unzip the package and launch the set up executable file Follow the installation wizard instructions, until the procedure is completed DT54xx LabView Control Software will be listed in the application menu Control Software operation Launch the DT54xx LabView Control Software, the following Graphical User Interface will open: Fig. 15: DT54xx Graphical User Interface Select the used port (VISA Port Select on the “Connection Box”) and click “Connect”; now the unit is ...
CAEN Electronic Instrumentation Fig. 17: Settings Box The “Power Enable Box” allows to power on/off the unit; the HV switch on the unit must be set on “Enable”, to do this. Fig. 18: Power Enable Box The Monitored values are reported by the relevant fields: Fig. 19: Monitor Fields The “Channel Status” box reports the actions that are going on: Fig. 20: Channel Status box The CLEAR ALARM button performs a software Alarm condition removal: Fig. 21: CLEAR ALARM button In order to disconnect from the Unit, click the relevant button on the Connection Box. DT54xx USB High Voltage Power Supply ...
Page 40
CAEN Electronic Instrumentation While the DT54xx LabView Control Software is running, a datalog.txt file is created in the sub folder CAEN_DT54XX_ControlSoftware of the default User’s destination folder; datalog.txt records all the changes of the parameters running. This function can be disabled, by writing the text disable into the datalog.txt file and saving it. DT54xx USB High Voltage Power Supply ...
Page 41
CAEN SpA is acknowledged as the only company in the world providing a complete range of High/Low Voltage Power Supply systems and Front‐End/Data Acquisition modules which meet IEEE Standards for Nuclear and Particle Physics. Extensive Research and Development capabilities have allowed CAEN SpA to play an important, long term role in this field. Our activities have always been at the forefront of technology, thanks to years of intensive collaborations with the most important Research Centres of the world. Our products appeal to a wide range of customers including engineers, scientists ...