Thermo Scientific 1300 Series Operating Manual page 51

Class ii, type a2 biological safety cabinet
Hide thumbs Also See for 1300 Series:
Table of Contents

Advertisement

Section 7
Specifications
Volume, weights, and loads
Description
Volume
Drain pan
Weights
Cabinet
Stand
Loads
Maximum load on work surface
Maximum load on overall working area
Ambient conditions
Description
Temperature
Max. ambient temp during operation
Min. ambient temp during operation
Humidity
Max. humidity / operation
Max. humidity / storage
Heat dissipation to environment
Room temperature 20°C
Room temperature rise
Above room temp with window closed
Ergonomics
Noise level*
*The noise level was determined in accordance with NSF/ANSI 49. 2004. The measurement uncertainty is within a range of ± 2 dB.
7-2
1300 Series A2
4 ft Models 1355, 1357, 1385, 1387
l / gal
kg / lb
kg / lb
kg / lb
kg / lb
4 ft Models 1355, 1357, 1385, 1387
°C / °F
°C / °F
% r.H.
% r.H.
kJ/s / BTU
°K
dB(A)
20 / 5.3
170 / 375
30 / 66
25 / 55
50 / 110
40 / 104
10 / 50
90, non-condensing
95
0.15 / 0.14
< 2
63, ±2
6 ft Models 1356, 1358, 1386, 1388
30 / 7.9
230 / 507
35 / 77
75 / 165
6 ft Models 1356, 1358, 1386, 1388
0.26 / 0.25
< 2
65, ±2
Thermo Scientific

Advertisement

Table of Contents
loading

This manual is also suitable for:

135513561357135813851386 ... Show all

Table of Contents