Error Messages - Fora Diamind VOICE Owner's Manual

Blood glucose monitoring system
Table of Contents

Advertisement

Available languages

Available languages

Error Messages

Message
What does
meter say?
Only Error
Beep
Only Error
Beep
Only Error
Beep
Only Error
Beep
Only Error
Beep
Only Error
Beep
Cause
There is insufficient
power in the batteries
to perform the test.
The strip has been
used.
The blood sample has
been applied to the
strip before the meter
was ready for the
measurement.
Ambient temperature
is below the operating
temperature.
Ambient temperature
is above the operating
temperature.
You may have
removed the strip
after applying blood,
or there is insufficient
blood volume.
What To Do
Replace the batteries
immediately and reset
the date and time on the
meter setting.
Repeat the test with a
new strip.
Repeat the test with a
new strip. Apply the blood
sample when the meter
indicates that it is ready
for the measurement
by displaying a "
and a "
" which starts
blinking.
The operating
temperature range is
from 10 ° C to 40 ° C (50 ° F
to 104 ° F). Repeat the
test after the device and
test strip have reached
the above-mentioned
operating temperature.
Review the instructions
and repeat the test with
a new test strip. Please
make sure you use the
correct technique and
dose with enough blood.
"
33