Error Messages

MESSAGE
Err 1
Err 2~Err 8
Err 10~Err 20
E-U
E-F
E-t
52
CAUSE
Problem with the cuff
or meter.
Problem with the
measurement.
Problem with the meter.
Strip has been used.
Test strip is removed
while counting down,
or insufficient blood
volume.
Environmental
temperature is
outside the system's
operational range.
Battery is low.
WHAT TO DO
Please contact local
customer service for help.
Refit cuff tightly and
correctly. Relax for 5
minutes and repeat the
measurement. If the
error still remains, please
contact the local cus-
tomer service for help.
Please contact the local
customer service for
assistance.
Repeat the test with a
new strip.
Review the instructions
and repeat the test with a
new strip. If problem
persists, contact the local
customer service for
assistance.
System operational range
is 10°C to 40°C (50°F to
104°F). Repeat the test
after the monitor and test
strip have reached the
above temperature
Recharge the battery
immediately.
.