HP 4262A Operating And Service Manual page 13

Lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Model 4262 A
1kHz
and 10kHz and about 900 milliseconds at
120Hz. The wide range capability of the 4262A
enables
a measurement
range
from
small capaci-
tances
such
as
mica
capacitors
and
the
para-
sitic
capacitance
of
a
semiconductor
device
through high capacitances such as the measurement
of electrolytic capacitors to be covered. A wide
range
of inductance measurements
from the in-
ductance of a high frequency transformer to that
of a power transformer can be measured. The wide
resistance range permits the measurement of wire-
wound resistors through the measurement of solid
resistors.
In parallel
capacitance
measurements,
either
a
test signal level of 1Vrms, or 50mVrms
can be selected.
1-7. The
4262A
has
the
capability
of making
..capacitance,
inductance, and resistance deviation...
measurements, This function is enabled by pushing
the A LCR switch to display the deviation of a re-
ference value. When the A LCR switch is depressed
the reference
value
is obtained
and memorized
from
the
preceding
measurement.
The practi-
cal use of this feature is evident when it is desired
to make a measurement on a variable capacitor:
First, the minimum
value is measured, then the A
LCR button is pushed. Minimum to maximum capa-
citance is now displayed as the capacitor is rotated
through its range. For parallel capacitance meas-
urements,
test signal levels of either 1Vrms
or
50mVrms may be selected. Other versatile 4262A
capabilities and features are, for example, the use
of internal and external de bias voltages, LC
zero
adjustment,
and options providing BCD
output,
HP-IB interfacing capability, or a comparator func-
tion.
1-8. SPECIFICATIONS.
1-9. Complete specifications of the Model 4262A
LCR Meter are given in Table 1-1. These specifi-
cations
are the performance standards
or limits
against which
the instrument is tested. The test
procedures
for the specifications are covered in
Section
IV Performance Tests. Table 1-2 lists gen-
MVOKEGAWA-HEMIETT -PARALDO °
Oe
HEPA
Section I
Paragraphs 1-7 to 1-16
1-10.
SAFETY CONSIDERATIONS.
MANUAL.
1-15.
An
instrument
manufactured
after
the
printing of this manual may have a serial number
prefix that is not listed on the title page. This un-
listed serial number prefix indicates the instrument
is different from those described in this manual.
The manual
for this new
instrument may
be ac-
companied
by a yellow Manual Changes supple-
ment or have
a different manual
part
number.
This supplement
contains
"change information"
that explains how
to adapt the manual
to the
newer instrument.
1-7

Advertisement

Table of Contents
loading

Table of Contents