HP 4262A Operating And Service Manual page 39

Lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Model 4262 A
Section II
Figure 3-6
MEASUREMENT
PROCEDURE
FOR
GENERAL
COMPONENTS
DC
POWER
SUPPLY
1. Remove shorting bar connections between high terminals and between low terminals (all
terminals are now isolated from each other). Connect 16061A Test Fixture to 4262A
UNKNOWN terminals.
Note
User constructed test fixture may also be
connected.
Guard terminal is sometimes used
in small capacitance measurements.
2.
Depress LINE button to turn instrument on. An initial display test is automatically per-
formed before measurement begins.
3. Check that 4262A trigger lamp begins to flash. The 4262A control functions are auto-
matically set as follows (automatic initial settings):
DC BIAS
ns
ae
ee
ae
OFF
SELF TEST
Rm
eg ea
OFF
CIRCUIT MODE. „u...
AUTO
FUNCTION:
use
ae ae
C
LCR RANGE
oode cene
a
e RS
AUTO
FOSS
oe
ra
ac he
Zul
a
haa Rs S
D
DO RANGE
hi. 545 EE een
AUTO
TEST SIGNAL sinne
ikHz
TRIGGER
ra
ard nenn er INT
Note
To check fundamental operating conditions of
the instrument, perform SELF TEST (refer to
Paragraph 3-5 for SELF TEST details). Press
SELF TEST button again to release the function.
4. Rotate C ZERO ADJ control until capacitance readout is 000 counts on LCR DISPLAY
(minus sign should not appear).
5. Connect a shorting lead to Test Fixture to short-circuit the Unknown terminals to zero
ohms (zero microhenries).
6.
Press
L FUNCTION button.
Figure 3-7. General Component Measurements (Sheet 1 of 3).