HP 4262A Operating And Service Manual page 60

Lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Section II
Paragraphs 3-71 to 3-75
3-71.
Remote/Local Capability: RL1.
The
4262A
goes to Remote
Status only when it
accepts Listen address with REN (Remote Enable)
line in the Bus lines set to "1". Remote status is
not obtained if REN line is set to "1" after Listen
address is received. Remote status is returned to
Local status when
one
of following conditions is
present:
(i)
REN line is set to "0".
(2)
LOCAL switch on front panel is depressed.
(3)
GTL (Go To Local) command is received.
Local Lock Out: LLO
3-72.
Device Clear Capability: DC1.
3-73.
Device Trigger Capability: DT1.
A5 A4 A3 A2 Al
0
0
0
0
OO
eee
0
Model 4262A