Page 1
Operation Manual BCFHR-303 Laboratory High Speed Refrigerated Centrifuge Thank you for Choosing Biolab products. Please read the “Operating Instructions” and “Warranty” before operating this unit to assure proper operation.
Page 2
Laboratory High Speed Refrigerated Centrifuge We appreciate your business with Biolab. To prevent any potential accident, please operate centrifuges according to the following safety protocol. 1. Unplug the main power cord, when performing maintenance or when centrifuge is expected not being used for a long period of time.
Page 4
Laboratory High Speed Refrigerated Centrifuge 01Introduction: BCFHR-303 High speed refrigerated centrifuge is an ideal centrifuge for blood bank, hospital and laboratory. Main specification and technical parameters: Max. Speed 21000 r/min Max. Volume 4×1000ml Max. RCF 46140×g Temperature range -20℃ ~ 40℃...
Page 6
Laboratory High Speed Refrigerated Centrifuge 03 Operation: 1.Checking whether the round nut in the rotor is loose, if so, fasten them. 2.Prepare the tubes; put the test liquids into the tubes, checking whether they are in the same level with your eyes. 3.Load tubes on the tube rack symmetrically;...
Page 7
Vertical Autoclave BAVT 200 C) Speed window This window can set up two kinds of content “speed and RCF×g by pressing the button “RCF” to switch each other. What is more, to press the buttons “▲ and ▼”to increase or reduce the figure.Do not exceed the Max.speed &...
Page 8
Laboratory High Speed Refrigerated Centrifuge Figure 2. ) button and ▼ button ▲ Keep touching ▲ button , the number will continuously increase , Click ▲ button , the number will increase bit -by -bit ; Keep touching ▼ button, not stop , the number will continuously reduce , Click ▼ button , the number will reduce bit-by-bit.
Page 9
Vertical Autoclave BAVT 200 • Stop the centrifuge When the time window count down to 0 ( or press the stop button ), stop indicator will be brighten, the speed will be down to 0, then the centrifuge will stop running. •...
Page 10
Laboratory High Speed Refrigerated Centrifuge please unplug the main power cord, so the centrifuge can cut off all the power supply. 6.In order to protect the safety of operator, cannot allow to open the lid when the instrument running. Instrument itself is also with this kind of protection function, that is to say, when speed is over the max speed with 100rpm, the lid switch will cut off the power automatically.
Page 11
Vertical Autoclave BAVT 200 Trouble-shooting service Trouble Cause Removal No 220/110V power The display window is not Check the power supply supply working when power is connected Fuse is broken Check and replace fuse The connector to circuit board connect it properly is not properly connected.